Design of Multiplex Lateral Flow Tests: A Case Study for Simultaneous Detection of Three Antibiotics

Submitted:

31 December 2019

Posted:

31 December 2019

You are already at the latest version

A peer-reviewed article of this preprint also exists.

Abstract
The presented study is focused on the impact of binding zones locations at immunochromatographic test strips into analytical parameters of multiplex lateral flow assay. Due to non-equilibrium conditions for such assays the duration of immune reactions influences significantly on analytical parameters, and the integration of several analytes into one multiplex strip may cause essential decrease of sensitivity. To choose the best location of binding zones, we have tested reactants for immunochromatographic assays of lincomycin, chloramphenicol, and tetracycline. The influence of the distance to the binding zones on the intensity of coloration and limit of detection (LOD) was rather different. Basing on the obtained data, the best order of binding zones was chosen. In comparison with non-optimal location the LODs were 5-10 fold improved. The final assay provides LODs 0.4, 0.4 and 1.0 ng/mL for lincomycin, chloramphenicol, and tetracycline, respectively. The proposed approach can be applied for multiassays of other analytes.
Keywords: 
;  ;  ;  ;  
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.

Downloads

403

Views

608

Comments

0

Subscription

Notify me about updates to this article or when a peer-reviewed version is published.

Email

Prerpints.org logo

Preprints.org is a free preprint server supported by MDPI in Basel, Switzerland.

Subscribe

© 2025 MDPI (Basel, Switzerland) unless otherwise stated