1. Introduction
In past years recreational diving has attracted more and more interest[1] and among such divers most likely an increasing number of divers living with diabetes (as reflected in the research cited throughout this brief report [
1]). At the same time technological advances such as increased use of continuous glucose monitoring (CGM) not only improve therapy but also could increase safety of people living with diabetes and those around them [
2].
Examples of studies
validating CGM before and after dive, Bonomo et al. [
3] indicated during „Deep Monitoring Programme“ that Medtronic CGM delivered plausible values during diving; similarly, Lormeau et al. [
4] indicated that Freestyle Libre may run properly also under pressure/after recompression. Similarly, Adolfsson et al. [
5] tested Medtronic CGM on 117 dives. As early as 2004 Dear et al. [
6] surveyed plasma glucose of 555 dives. Pollock et al. [
7] monitored plasma glucose of teenage divers before and after dives.
Studies
testing CGM sensors in compression chambers comprise inter alia Adolfsson et al. [
8] who tested 48 enlite sensors in hypo- and hyperbaric conditions up to 4bar additional pressure (where MARD even decreased under hyperbaric conditions) or Pieri et al [
9] who showed that Dexcom G4 work accurately under hyperbaric conditions, collecting 26 data points. Both studies were conducted in compression chambers.
The use and validation of CGM results is an emerging theme in diving and research. However, there is limited evidence under hyperbaric conditions for individual systems and so far, no study was conducted under real life scuba diving conditions. This may also be one reason why diving guidelines[2] make little reference to the use of CGM to increase the safety of divers living with diabetes. This research aims to explore usability of CGM systems during diving and derive implications for divers living with diabetes.
In continuation of the above, this
exploratory study aims to add to such data for validation and testing contemporary CGM systems. Further, it aims to provide evidence on the accuracy of CGM systems
during real life scuba use including exposure to both pressure and salt water. In order to do so, four dives were conducted with one subject during which blood glucose values were collected to then be compared against CGM readings and test MARD values [
10] and Clarke Error Grid Distirbution [
11].
2. Experimental design
The experiment was set up to enable blood glucose measurement underwater, i.e., under hyperbaric conditions with up to 2 bar additional pressure. All four dives were conducted in the open sea in Tolo, Greece during August 2022 with the support and under the guidance of the dive center Intro Dive, Tolo. Temperature was ca. >30°C air, ca. 25°C in water at surface and further ca. 2°C lower per 10m depth. To enable blood glucose measurement underwater, a dry bag was used to create a dry bubble: the dry bag was used upside-down filled with air and pulled down by a weight system to establish neutral buoyancy. Inside the drybag a smaller bag was mounted to carry the blood glucose measurement equipment as well as a towel to dry off subject’s fingers as good as possible. During the dives the subject would bring in its hand from underneath into the air space inside the dry bag. In there fingers would be toweled off and blood glucose sample taken. Below two pictures of the setup and carrying under water:
The exploratory experiment was conducted with one subject: male, T1D for 5 years, applying conventional pump therapy for 3 years, with a HbA1C of <=6%, and time in range of >90% in the four quarters preceding the experiment. The subject reported no relevant hypoglycemic events during the week of the experiment. To note that ca. 1h before the third dive a hypoglycemia occurred with sensor readings as low as 73mg/dl (Dexcom G6) and 65mg/dl (Libre 3) vs. BG measurements of 73/77mg/dl (cf. figure 5). Other than this event, IG was in a range of 70-165mg/dl in the time 2h before the dive start, throughout the dives, and 1h after the dive end. Slope was 1mg/dl/min (except for 2 samples at up to 4mg/dl/min Dexcom / 3 samples at up 3mg/dl/min Libre); i.e., minimal distortions expected.
The subject disconnected the pump during preparations for the dives ca. 0.5-1h prior to the start of the dive which led c.p. to a positive glucose trend during the dives. The subject corrected these after the dives based on own assessment (see dive logs in annex for details). Considering the range of glucose values renal function was not monitored. However, it is worth noting for divers living with diabetes that a different management and monitoring may be more suitable depending on their individual profile and therapy.
As the subject conducted the blood glucose measurement under water itself with a familiar device, only some preparatory briefing was provided by diabetologist and the guiding divemaster.
The following equipment was used during the experiment:
The dive profiles were recorded via a Cressi Newton dive computer which recorded measured depth every 20 seconds.
The blood sugar values were collected using mylife Unio Neva[3].
Two CGM systems were tested:
Dexcom G6 sensor inserted four days before the first dive and calibrations with BG stabilized two days before the first dive. The sensor expired 2 days after the last dive. It was worn on the abdomen in line with manufacturer guidance.
Abbott Freestyle Libre 3 sensor inserted 3 days before the first dive and expired 7 days after the last dive. It was worn on the upper arm in line with manufacturer guidance.
The CGM sensors were positioned on the abdomen of the subject. They were worn under the wetsuit but without further (environmental) protection/fixation.
In general, every sample would comprise two blood glucose measurements to reduce impact form inaccuracies in the blood glucose values. The sampling procedure was repeated every 5 minutes or longer, corresponding to the frequency of the interstitial glucose recordings via the CGM systems. In total 33 samples were collected, of which 26 under water at various depth/relative pressure, 4 before diving, and 3 after diving. One post dive sample could not be collected as BG reader dropped into the water during exit.
Annex shows the results for each of the dives over time including 2h before and 1h after the dives. Note that none of the BG samples collected was used to calibrate the CGM because otherwise CGM values could have been affected by (re-)adjustment effects.
3. Results
The samples of blood glucose collected during the dive were mapped to the corresponding CGM reading and the absolute relative deviation was calculated to track the accuracy of CGM readings vs. actual blood glucose. The table below shows the mean absolute relative deviation (MARD) of the samples pre and post dive as well as during the dive (clustered by depth/relative pressure) and compares these against the MARD as quoted by the manufacturer under normal conditions (first column). MARD was chosen as measure to be consistent and comparable with the measure of accuracy published by the manufacturers:
Generally, two BG samples were taken and the average of the two was used to calculate MARD in order to minimize impact from inaccuracies of BG measurements. Where the dive time realized under water that the value was implausible, e.g., because diluted, a new sample was taken immediately (e.g., „low“ or in the 40ies despite being above 100 shortly before). In one case such resampling took too long and the two samples were used as individual data points as the CGM value updated in the meantime. In three further cases one of the two blood samples was discarded during analysis of the data because they were obviously implausible as they showed both, a larger deviation and off the trend of neighboring samples.
In the 28 cases where the average of two blood samples was used the median spread around such average was 2.3% (3.7% mean). Even adding this to the observed relative deviations does not indicate undue misreading by CGM under pressure/after decompression.
Further, no correlation between depth and observed deviation between BG and IG could be observed as per exhibit 2 below. However, it is important to note the limitations of the small sample. In order to verify this first impression, increased sample size and regression controlling for other factors like BG level would be required.
Figure 1.
Absolute relative deviation vs. depth. Source: experimental dives.
Figure 1.
Absolute relative deviation vs. depth. Source: experimental dives.
As can be seen from the dive charts in annex, no systematic pattern of deviation emerged between blood glucose and tissue glucose. Comparing the blood glucose samples during the dives to those pre/post, it appears the samples were not diluted by seawater.
For Dexcom 25% of samples turned out above the CGM reading leading to the suspicion that if they/some of them were used for calibration, accuracy of results may even have improved. For Libre 42% of samples turned out below the CGM reading, i.e., appears balanced.
All samples were taken while CGM values in range of 70-145 Dexcom / 60-150 Libre and slope was 1mg/dl/min (except for 2 samples at up to 4mg/dl/min Dexcom / 3 samples at up 3mg/dl/min Libre); i.e., minimal distortions expected.
Considering the time lag between CGM readings (based tissue glucose) and actual blood glucose, one could have expected this effect emerging in the samples shortly after a pressure increase/decrease (i.e., decent/ascent). However, no such pattern emerged as can be seen from the dive logs in annex. Similarly, (abrupt) temperature changes/temperature level may have an effect on tissue glucose and/or CGM readings as already indicated by Adolfsson et al.8. No pattern emerged in this setting with temperature decreased a) from air to water (from ca. >30°C to ca. 25°C) and b) in deeper waters (ca. 2°C per 10m).
According to the International Organization for Standardization (ISO guideline 15197), a maximum acceptable difference within 20% for glucose levels > 75mg/dL and within 15mg/dL for levels < 75mg/dL in at least 95% of the values compared. The chart below shows that this criterion is fulfilled. All but one reading of the Dexcom fall into area „A“ of the Clarke Error Grid. The only outlier is a reading at ca. 17m depth; the subsequent readings at the same depth showed 6-12% deviation. Furthermore, all readings of the Libre 3 fall into area „A“ of the Clarke Error Grid.
Figure 2.
MARD under hyperbaric conditions vs. Manufacturer data. Source: experimental dives, Clarke et al. (1987).
Figure 2.
MARD under hyperbaric conditions vs. Manufacturer data. Source: experimental dives, Clarke et al. (1987).
An obviously critical point is the rather small sample size (33 in total, of which 4 pre-dive, 26 under hyperbaric conditions, 3 post-dive). To note that similar conclusions were reached based on a sample of 26 in a hyperbaric chamber using Dexcom G49 and 48 enlite sensors up to 4bar additional pressure in a chamber8. Whilst the results cannot be added together because different systems were tested, they are consistent in the sense encouraging divers living with diabetes to carefully consider of CGM data bearing in mind limitations in manufacturers‘ clearances.
4. Conclusion
For practitioners the result of this study gives a first indication that Dexcom G6/Libre 3 CGM can be used in planning and monitoring dives in order to increase divers‘ safety. However, further testing and evidence should be collected before issuing a recommendation as currently CGM certified by manufacturer only for up to e.g., 2.5m for 24h[4] / 1m for 0.5h[5]. Once the results could be confirmed it seems advisable to update safety-recommendations to divers living with diabetes to take account of CGM technology:
Dives can be planned based on a combination of CGM observed trend and own extrapolation based on basic indicators such as active insulin, recent and expected activity level, carbohydrate intake, and typical trends that time of the day [
12]. Especially Moser et al. [
13] have developed detailed recommendations for protocols to manage blood sugar between active insulin and exercise. CGM curve up to beginning of dive offers more insight than insulated blood glucose values (cf. examples in annex)
Divers living with diabetes may consider taking CGM-sensor and –monitor under water to increase safety. In particular, monitoring glucose values enables underwater intake of glucose: Some of the current guidelines recommend carrying carbohydrate supplies under water. However, taking out regulator to supplement sugar when sensing hypoglycemia can be dangerous symptoms include impaired motion control and consciousness. Using a CGM monitor for early monitoring allows to supplement sugar before actual hypoglycemic event.
This is particularly viable as the use of electronics during dives appears to be increasing. This includes phones in protective cases used as camera – the same phone that may have the CGM monitor app installed. One area of concern may be Bluetooth connectivity through water, especially at greater depth. Workarounds to be explored could include bringing the receiver close to the sensor (possibly even under the exposure suit) and a) only take out for checks or b) rely on vibration and/or acoustic alerts (to be confirmed they would be recognized under real life diving conditions). Whilst research can provide supporting evidence of possible methods every diver will have to test which setup works best in practice considering individual gear and personal preferences / circumstances.
Organizations issuing guidelines for safe diving are encouraged to weigh the benefits of using CGM at least for indicative monitoring against the fact that this exceeds the range of use warranted by manufacturers. The experimental setup developed here appears a cost-effective way to collect further evidence and test also other CGM models. Both divers and organizations should bear in mind that this exploratory study was limited to one subject under controlled conditions, reasonably warm water, and with good visibility. As part of further research additional data points should be gathered and including alterations like dry suit to corroborate the case for use of CGM during recreational scuba diving.
For academia the experiment achieved milestones in terms of being the first study to a) overlay CGM data with dive profile and b) collect blood glucose values under water (real life situation) to verify CGM values. The results add to the research on the accuracy of CGM systems during dives quoted in the introduction; In particular to the validations under hyperbaric conditions8,9.The results indicate for two further CGM systems that CGM readings in hyperbaric situations and after decompression are not less accurate than under normal use. Notably, the experimental setup allowed testing in real life conditions under water in an efficient manner which can be considered a step beyond compression chamber testing.
In terms of further research there is a strong need to confirm this exploratory research by increasing size of sample and collecting samples under even higher additional pressure (recreational diving up to 40m implies up to 4bar additional pressure; so far only tested in compression chamber8). As a variation to the direct BG testing under water with handheld devices, one may consider collecting blood samples for more accurate laboratory analysis to increase precision of the results.
Further analysis could be conducted, such as testing for impact of temperature (and interplay with exposure protection, e.g., divers wearing dry suit or lycra only), use of oxygen enriched air [
14], or activity level (e.g. using heart rate monitor to track strain). From the literature, temperature appears to be the most relevant factor: Adolfsson et al. [
8] already refer to the impact of temperature on CGM readings[6]. A systematic realistic approach would be to conduct experiments in different water temperatures from ice diving to tropical diving.
An increased sample size would further enable more thorough statistical analysis – including whether depth or other factors during diving have an impact on CGM accuracy. For additional certainty, one may consider collecting blood samples under water which then are tested for BG a) at normobaric conditions and b) using higher precision BG measurement devices.
The experimental setup developed to collect blood glucose samples under water proved to be an efficient option for future data collection, be it to confirm results on systems tested to date or to test new CGM systems as they come to market.
Annex: Dive logs
Figure 3.
Dive log #1. Source: experimental dives.
Figure 3.
Dive log #1. Source: experimental dives.
Active insulin: pump disconnected ca. 3h before dive; minimal tail from correction bolus
Sport: none 2h before
Food: none 2h before
Expectation: no exercise, no food, no insulin > steady to rising
In hindsight: confirmed; rise started when i) all insulin depleted and ii) freezing began
Note: no post-dive blood glucose measurement because of technical issue
Figure 4.
dive log #2. Source: experimental dives.
Figure 4.
dive log #2. Source: experimental dives.
Active insulin: pump disconnected 50‘ before dive; minimal correction bolus at that moment
Sport: none 2h before
Food: none 2h before
Expectation: stable as no food and no exercise; no basal, but remaining bolus
In hindsight: confirmed
Figure 5.
dive log #3. Source: experimental dives.
Figure 5.
dive log #3. Source: experimental dives.
Active insulin: pump disconnected 45‘ before dive; small correction bolus 2.5h before dive
Sport: none 2h before
Food: 2KE 50‘ before dive
Expectation: ?: Remaining bolus vs possible late digestion of food; no exercise and no basal
In hindsight: stable
Figure 6.
dive log #4. Source: experimental dives.
Figure 6.
dive log #4. Source: experimental dives.
Active insulin: pump disconnected 1h pre-dive; no active bolus
Sport: none 2h before
Food: none 2h before
Expectation: increasing: no active insulin, no sport, no food
In hindsight: confirmed
Author Contributions
Conceptualization: Both; Methodology: Both; ValidationPrimarily Daniel; Formal
Analysis Primarily Philipp; Resources: both; Data curation, Writing - Original Draft Preparation, Visualization:
Philipp; Writing – Review & Editing: Daniel
Ethical review and approval
not applicable for this brief report: only analyzing BG/IG values that subjects collect
as part of their ongoing monitoring anyway. Subject is diving with CGM sensors anyhow so that data collection
for this brief report did not create additional risk and honors the pricniples of the Helsinki Declaration.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.
Acknowledgements
This study would not have been possible without the diving equipment and logisitcal
support from Intro Dive, Tolo. A particular thanks to divemaster Will van den Heuvel for his support during
the dives.
Conflicts of interest
The authors declare that they have no conflicts of interest.
[2] E.g.:
Globally: https://dan.org/health-medicine/health-resource/health-safety-guidelines/guidelinesfor-
diabetes-and-recreational-diving/ [15] and https://dan.org/safety-prevention/diversafety/
divers-blog/scuba-diving-and-diabetes/ [16]
https://pros-blog.padi.com/diving-and-diabetes/
Italy2
Australia: https://diabetessociety.com.au/documents/ADS_Diving_Diabetes_2016_Final.pdf
[3] Before conducting the experimental dives it was not clear whether pressure may have an effect on the functioning of the device, e.g. because of blood denisty on sensor. Considering the plausibility of the results, also vis a vis pre/post dive measurments, it is assumed that the BG device was not affected by pressure.
[6] Similarly, the study participant reported that when swimming without exposure suit a steep drop in the level of measured IG is observed when entering the water; which could be due to either the sensor being affected technically and/or a change in glucsoe level in the interstitium under the skin.
References
- Edge et al. (2005): Edge CJ, St Leger Dowse M, Bryson P. Scuba diving with diabetes mellitus--the UK experience 1991-2001. Undersea Hyperb Med. 2005 Jan-Feb;32(1):27-37. PMID: 15796312.
- Bonomo et al. (2019): Matteo Andrea Bonomo, Giovanni Careddu, Gerardo Corigliano, Paolo Di Bartolo, Pasquale Longobardi, Andrea Fazi, Elena Cimino, Elena Gamarra, Umberto Valentini: La persona con diabete tipo 1 e le immersioni subacquee. Rassegna Vol. 31, N. 1, marzo 2019: 22-40.
- Bonomo et al. (2009): Bonomo M, Cairoli R, Verde G, Morelli L, Moreo A, Delle Grottaglie M, Brambilla MC, Meneghini E, Aghemo P, Corigliano G, Marroni A. Safety of recreational scuba diving in type 1 diabetic patients: the Deep Monitoring programme. Diabetes Metab 35: 101-107, 2009.
- Lormeau et al. (2019): B. Lormeau, S. Pichat, L. Dufaitre, A. Chamouine, M. Gataa, J. Rastami, C. Coll-Lormeau, G. Goury, A.-L. François, V. Etien, J.-L. Blanchard, D. Hervé, A. Sola-Gazagnes, Impact of a sports project centered on scuba diving for adolescents with type 1 diabetes mellitus: New guidelines for adolescent recreational diving, a modification of the French regulations, Archives de Pédiatrie, Volume 26, Issue 3,2019, Pages 161-167, ISSN 0929-693X. [CrossRef]
- Adolfsson et al. (2008): Adolfsson P, Ornhagen H, Jendle J. The benefits of continuous glucose monitoring and a glucose monitoring schedule in individuals with type 1 diabetes during recreational diving. J Diabetes Sci Technol. 2008 Sep;2(5):778-84. [CrossRef]
- Dear et al. (2004): Dear Gde L, Pollock NW, Uguccioni DM, Dovenbarger J, Feinglos MN, Moon RE. Plasma glucose responses in recreational divers with insulin-requiring diabetes. Undersea Hyperb Med. 2004 Fall;31(3):291-301. PMID: 15568417.
- Pollock et al. (2005): Pollock NW, Uguccioni DM, Dear GdeL, eds. Diabetes and recreational diving: guidelines for the future. Proceedings of the UHMS/DAN 2005 June 19 Workshop. Durham, NC: Divers Alert Network; 2005.
- Adolfsson et al. (2012): Adolfsson P, Örnhagen H, Eriksson BM, Cooper K, Jendle J. Continuous glucose monitoring--a study of the Enlite sensor during hypo- and hyperbaric conditions. Diabetes Technol Ther. 2012 Jun;14(6):527-32. [CrossRef] [PubMed]
- Pieri et al. (2016): “Continuous real time monitoring and recording of glycaemia during scuba diving: pilot study” – Pieri M, Cialoni D, Marroni A, Undersea Hyperb Med. 2016 May-Jun; 43(3):265-72.
- Reiter et al. (2017): Reiterer F, Polterauer P, Schoemaker M, et al. Significance and Reliability of MARD for the Accuracy of CGM Systems. Journal of Diabetes Science and Technology. 2017;11(1):59-67. [CrossRef]
- Clarke et al (1987): Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL: Evaluating clinical accuracy of systems for selfmonitoring of blood glucose. Diabetes Care 1987; 10: 622-628.
- Johnson (2016): Johnson R. A day in the life of a diabetic diver: the Undersea and Hyperbaric Medical Society/Divers Alert Network protocol for diving with diabetes in action. Diving Hyperb Med. 2016 Sep;46(3):181-185. PMID: 27723021.
- Moser et al (2020): Glucose management for exercise using continuous glucose monitoring (CGM) and intermittently scanned CGM (isCGM) systems in type 1 diabetes. Diabetologia,. [CrossRef]
- Lormeau et al. (2006): B. Lormeau , A. Sola-Gazagnes , L. Dufaitre , A. Tabah , O. Thurninger, N. Assad , P. Valensi: Intérêt d’un mélange suroxygéné (Nitrox) en plongée sous marine chez le patient diabétique de type 1. Diabetes & Metabolism 2006. [CrossRef]
- Divers Alert Network 2020: Guidelines for Diabetes and Recreational Diving - Proceedings Summary | DAN/UHMS Diabetes and Recreational Diving Workshop, https://dan.org/wp-content/uploads/2020/07/diabetes_proceedingsummary_final.pdf.
- Pollock et al. (2006): Pollock NW, Uguccioni DM, Dear Gd, Bates S, Albushies TM, Prosterman SA. Plasma glucose response to recreational diving in novice teenage divers with insulin-requiring diabetes mellitus. Undersea Hyperb Med. 2006 Mar-Apr;33(2):125-33. [PubMed]
|
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).