In the analysis of flavorings and volatile organic compounds (VOC) in general, gas chromatography (GC) is the method of choice for separation of a broad spectrum of VOC. There are numerous GC-based applications for the analysis of complex matrices such as essential oils, flavors, or fragrances, particularly in combination with mass spectrometry (MS). However, in the last decade, ion mobility spectrometry (IMS) has gained increasing attention for the analysis of food, beverages, and flavorings [
1,
2]. IMS systems in hyphenation with GC are mostly realized in the form of drift tube IMS systems (DTIMS), while only few systems base on differential mobility (DMS) or asymmetric field (AIMS/FAIMS) [
3,
4]. For the sake of simplicity, IMS is used in the context of DTIMS in the following. GC-IMS systems base on
3H,
63Ni, CD or UV ionization, whereas the ionization by
3H is among the most common, due to the low safety restrictions and the broad applicability [
3,
5,
6]. Such systems offer a remarkably high sensitivity for polar and medium polar compounds, due to the soft ionization, and operate at ambient pressure. This simplifies their use substantially in contrast to vacuum-based MS detectors, with regard to robustness, ease of use and cost of systems. While IMS detectors commonly outperform standard quadrupole MS detectors in terms of sensitivity, one major drawback is the lack of comprehensive and commonly accepted databases. Whereas such libraries are widely used for substance identification in MS, this is not the case for IMS, where substances are typically identified relative to reference substances, comparing retention and drift time. For complex matrices, this is time consuming, tedious, and limited by the need for the availability of reference substances. In a previously published study, hyphenation of IMS and QMS showed a beneficial effect for substance identification, in combination with a soft ionization of the IMS and subsequently, valuable data for chemometric evaluation. Brendel
et al. were able to show that a classification of selected citrus fruit juices is possible by both QMS and IMS data based on enrichment-free, static headspace GC-QMS-IMS, but match quality in MS databases was limited due to low QMS S/N ratios and the intense fragmentation in EI [
2]. Additionally, Schanzmann
et al. described the combination of thermal desorption with GC-MS-IMS for breath analysis. They were able to show a good accordance of MS and IMS retention times for a homologous series of ketones up to 2-decanone. Within the study, they were also able to identify five VOC, including ethanol, isoprene, acetone, 2-propanol, and 1-propanol, in breath samples with the NIST database [
7]. This approach is typically not applicable for complex flavorings and food samples, as these show a broad range of VOC, ranging from ketones, esters to terpenes and sesquiterpenes, with boiling points up to 260 °C. In nearly all cases, QMS sensitivity was the limiting factor. To overcome this limitation, and to obtain m/z data also for complex matrices, a new trapped-headspace (THS)-GC-MS-IMS setup was designed. This involved the optimization of retention gap dimensions and APC pressure settings, in order to generate reasonable signal intensities on both detectors. The main challenge here was the balance between sufficient S/N in QMS and not overloading the IMS detector. In
Figure 1, the THS-GC-MS-IMS setup and the flow directions are visualized. The hyphenation with the described trapped-headspace sampler is beneficial for sampling in static as well as dynamic headspace, in particular due to the constant control of the vial pressure and the ability of concentration steps.
1.1. Trapped Headspace Sampling
The basic principle for headspace sampling is the equilibration of analytes in sample solution and the vapor phase above [
8]. The partition coefficient, describing the ratio of the analyte concentration between the phases, is shown in Equation 1 [
9].Compounds with a low partition coefficient
K, will be more abundant in the vapor phase, while compounds with a high partition coefficient will be more prevalent in the sample phase. Besides the analyte’s vaporization enthalpy and vapor pressure, the solubility in sample phase is particularly important for the partition coefficient. Thus, in method development, vial temperature, adjustment of pH-value, and the addition of salt are effective and important parameters [
9].
where:
K = partition coefficient of a defined compounds
CS = concentration of the compound in sample phase
CG = concentration of the compound in vapor phase
For headspace sampling, the vapor phase of a substance is sampled and analyzed. The sample is sealed in a vial, heated and brought to an equilibrium [
8]. In static headspace sampling (SHS), a defined volume of the vapor phase is extracted using a syringe, resulting in analyte fractions in ppmv levels. This results in absolute amounts in the lower pg range on column. Paired with the system-immanent issue of high fragmentation behavior of sensitive compounds in EI mode and the limited sensitivity of QMS detectors in full scan, this poses a challenge for QMS based systems. Consequently, the majority of polar and medium polar analytes are detected mainly in IMS and only a limited number of these are detectable in the QMS detector. One elegant way around this limitation is trapped or dynamic headspace sampling (THS). It allows for a pre-concentration of the analytes up to sub-ppbv level [
9]. While for matrices with higher analyte concentrations, classical SHS is an effective, feasible option, THS excels at samples with low abundant analytes in complex matrices [
8]. This is of particular advantage in quality analytics of food and beverages, as the incubation temperature is limited due to the risk of transformation reactions and formation of artifacts.
In
Figure 2, the principles of THS sampling are visualized. In a first step, the sample is pressurized and the volatiles are continuously evaporated to the vapor phase [
10]. Subsequently, the vapor phase is trapped with a sorbent, extracting VOC from the gas phase [
8,
9]. The last step is the desorption of the trapped analytes and the transfer onto the GC column [
10]. In multiple headspace extraction (MHE) mode, these steps are repeated several times, ensuring that the majority of the volatiles are extracted from the sample phase [
9,
10]. In MHE, the sample is equilibrated again after each sampling procedure. The trap material, such as TenaxTA or carbon-based sorbents, must fulfill several requirements. The material should be capable to extract a broad range of volatile samples and retain all analytes of interest, further allow for a fast injection on the GC column and should have a minimal carry-over effect of impurities to the sample [
9]. For trapping, there are different approaches, such as cryogenic, electrically-cooled or with solid sorbents, as well as liquid films in solid support [
11]. Applications for THS sampling are found in a broad spectrum of analytical tasks in foods, feeds, pharmaceuticals, and environmental samples [
12,
13,
14,
15,
16].
1.2. Ion Mobility Spectrometry in VOC Analysis
Within the last decades, GC-hyphenated IMS gained increasing popularity for the analysis of VOC in all types of fields due to its high sensitivity in combination with a robust design. The applications range from process and quality control to explosives and drugs, foods, beverages, and flavor products [
2,
17,
18,
19]. The IMS cells operate at ambient pressure and
3H-based ionization of the analytes is a reaction of proton water clusters H
+[H
2O]
x with the analytes, forming protonated monomers MH
+[H
2O]
n-x as shown in Equation (2) and at higher analyte concentrations, dimers M
2H
+[H
2O]
n-x, as shown in Equation (3) [
6]. The proton water clusters are formed within a reaction cascade, initiated by the tritium source (< 100 MBq).
The formation of proton water clusters, also called reactant ions, depends on residual moisture content of the gas atmosphere [
3,
6]. After ionization, monomers and dimers are accelerated into the drift tube by a weak electrical field against a defined flow of a drift or buffer gas, typically nitrogen. Due to collisions with drift gas molecules, ions are decelerated according to their collision cross section (CCS). Thus, drift time is dependent on mass, charge and structure, resulting in a characteristic drift time for each analyte. The soft, APcI-like ionization leads to excellent sensitivity ranges of DTIMS instruments, which makes them optimal tools for non-target strategies (NTS). These approaches are based on the complete (amenable) spectral fingerprint instead of using defined marker compounds [
3].
In VOC profiling of foods and flavors, IMS is already used in a multitude of applications, ranging from quality analysis of olive oils, authenticity of honey, quality control of brewing hops, profiling of dairy products, such as kefir or for discrimination of different citrus juices [
1,
2,
20,
21,
22,
23].
1.3. VOC Analysis of Mangos
Mango fruits (
Mangifera Indica L.) are considered as the “King of Fruits” and particularly mango purees are of substantial value for food and beverage industry [
24]. There are more than 100 cultivars available, which differ in taste and flavor. In general, the flavor is described as fruity, sweet and floral, but some cultivars display different, citrus and terpene notes. Due to that flavor profile, the Indian cultivar ‘Alphonso’ is one of the most popular and shows an increasing demand on the world market [
25]. Although there are numerous studies on mangos and their volatiles, most of these focus on regional cultivars or on farm-grown fruits and not on partially processed commercial products, such as purees or pulps [
26,
27,
28]. Mango fruits display a complex volatile profile with more than 50 compounds, including alcohols, ketones, aldehydes, esters, monoterpenes and sesquiterpenes [
27,
29,
30,
31]. With regard to VOC composition, the cultivars can be distinguished into three groups. The first group features green, herbal flavors and 3-carene as the most abundant monoterpene, the second with rosin flavors and α-terpinolene as the major monoterpene compound and the last with (
Z)-β-ocimene as the predominant monoterpene, covering sweet, terpene and citrus notes, such as the cultivar ‘Alphonso’ [
32].
The majority of published studies on the VOC profiles of mango fruits and products hereof are target-based approaches and are dominated by SHS-GC-MS applications, mostly with solid phase microextraction (SPME) enrichment, Arrow, or other enrichment steps [
27,
33,
34,
35].
In previously published non-targeted studies, Tandel
et al. applied PCA and HCA to HS-SPME-GC-MS data and different extraction methods of Indian mango cultivars, while Farag
et al. employed PCA to analyze HS-SPME-GC-MS data of Egyptian mango cultivars [
27,
33]. Further, Shimizu
et al. reported on a volatile profiling of 17 different mango cultivars using HS-SPME-GC-MS in combination with PCA [
26].
So far, the literature on the use of GC-IMS in the context of mango fruits, purees, or other low-processed mango products is scarce. A number of authors reported on GC-IMS applications for post-harvest effects and focused on the terpene profiles in relation to fruit quality [
36,
37]. However, these studies focused specifically on Chinese cultivars and on the ripening process from green fruits to ripe fruits and used separate instruments for IMS and MS.
Consequently, the aim of the present study was to demonstrate the potential of THS sampling in combination with simultaneous GC-MS-IMS detection. While the study focused on the non-target approach to differentiate Mangifera indica L. cultivars, relevant metabolites were confirmed via MS in parallel to generate a more detailed insight. The prototypic system described here allows both approaches from one single injection. An additional source of information for the non-target approach is resulting from the use of a chiral GC column, which generates substantially more characteristic signals.