Preprint
Article

Semi-automatic Algorithms for Estimation and Tracking of AP-diameter of the IVC in Ultrasound Images

This version is not peer-reviewed.

Submitted:

29 October 2018

Posted:

30 October 2018

You are already at the latest version

A peer-reviewed article of this preprint also exists.

Abstract
Medical research has suggested that the anterior-posterior (AP)-diameter of the inferior vena cava (IVC) and its associated temporal variation as imaged by bedside ultrasound is useful in guiding fluid resuscitation of the critically-ill patients. This paper develops semi-automatic active ellipse and rectangle algorithms for measurement and tracking of the AP-diameter. The proposed algorithms are compared with an expert manual measurement and the previous work based on active circle model. It is shown that regardless of the shape of the IVC, the rectangle model always outperforms the two other models and performs very close to manual measurement.
Keywords: 
;  ;  ;  ;  ;  
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.

Downloads

299

Views

232

Comments

0

Subscription

Notify me about updates to this article or when a peer-reviewed version is published.

Email

Prerpints.org logo

Preprints.org is a free preprint server supported by MDPI in Basel, Switzerland.

Subscribe

© 2025 MDPI (Basel, Switzerland) unless otherwise stated