Preprint
Communication

Biomechanical Behavior Evaluation of a Novel Hybrid Occlusal Splint-mouthguard for Contact Sports: 3D-FEA

Altmetrics

Downloads

224

Views

191

Comments

0

A peer-reviewed article of this preprint also exists.

Submitted:

18 October 2021

Posted:

19 October 2021

You are already at the latest version

Alerts
Abstract
Orofacial injuries are common occurrences during contact sports activities; however, there is an absence of data regarding the performance of hybrid occlusal splint mouthguards, especially during compressive loading. To evaluate the total deformation and stress concentration, a skull model was selected and duplicated to receive two different designs of mouthguard devices: one model received a conventional custom-made mouthguard (MG) with 4-mm thickness and the other received a novel hybrid occlusal splint-mouthguard (HMG) with the same thickness. Both models were subdivided into finite elements. The frictionless contacts were used, and a nonlinear analysis was performed simulating the compressive loading in occlusion. The results were presented in von-Mises stress maps (MPa) and Total Deformation (mm). A higher stress concentration in teeth was observed for the model with the conventional MG, while the HMG design displayed a promising mechanical response with lower stress magnitude. The HMG de-sign displayed a higher magnitude of stress on its occlusal portion than the MG design. The hybrid mouthguard (HMG) reduced (1) jaw displacement during chewing and (2) the generated stresses in maxil-lary and mandibular teeth.
Keywords: 
Subject: Medicine and Pharmacology  -   Dentistry and Oral Surgery
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.
Prerpints.org logo

Preprints.org is a free preprint server supported by MDPI in Basel, Switzerland.

Subscribe

© 2024 MDPI (Basel, Switzerland) unless otherwise stated