Preprint
Article

Design and Manufacture of a Test Device for Small Vacuum Pumps

Altmetrics

Downloads

253

Views

169

Comments

0

A peer-reviewed article of this preprint also exists.

Submitted:

06 December 2022

Posted:

20 December 2022

You are already at the latest version

Alerts
Abstract
Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new or rebuilt pumps prior to installation, thus reducing downtime and radiation exposure during repairs. The Testbed incorporates a microprocessor that actuates the pump, valves, pressure sensor, and communicates with the user through lights, buttons, and an alphanumeric screen. The Testbed increases productivity and safety in the radiochemistry laboratory.
Keywords: 
Subject: Engineering  -   Industrial and Manufacturing Engineering
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.
Prerpints.org logo

Preprints.org is a free preprint server supported by MDPI in Basel, Switzerland.

Subscribe

© 2024 MDPI (Basel, Switzerland) unless otherwise stated