The PAS spectra shown in
Figure 5 depicts the test data containing unknown mixtures of water, ammonia and methane. As can be seen the mixtures with the three gas components has high spectral overlap. In the following we make a benchmark test of the relative accuracy of the PLS method compared with direct fitting with HITRAN spectra. The two main effects that can influence the performance and accuracy of the PLS method is high concentration of water relative to ammonia and methane concentrations and deviations in the wavelength axis compared to the training spectra. The four gas mixtures displayed in
Figure 5 are chosen as extreme cases to evaluate the performance of the PLS algorithm. The PLS method was again implemented using a standard training-test approach. The training data set was built starting from the three reference spectra for water, ammonia and methane, as shown in
Figure 2. The data set then expanded by means of calculating a linear combinations of 5000 reference spectra with superimposing Gaussian noise distributions. Based on the above calibration of the PLS method for the known concentrations we can estimate the concentrations of the unknown mixtures. We make the assumption that the water enhancement factor of ammonia has a similar behavior as methane.
Figure 5a shows data for a low concentration of water and relatively similar PAS signal strength for ammonia and methane with minimal deviation in the wavelength (less than 0.2 nm) axis compared to the training/HITRAN data. Keeping in mind that the Q-branch peak of methane is 10.4 times higher than the Q-branch of ammonia (as shown in
Figure 2). We find that the PLS method and HITRAN fitting agrees very well within the measurements uncertainties as shown in
Figure 6. The overall estimation of the water, ammonia and methane concentrations with the PLS method has a accuracy of 92%, 86% and 92%, respectively, relative to the HITRAN fitting. In
Figure 5b the water content is increased to 9000 ppm/V hereby diluting the concentration of ammonia and methane. In this case we find that the relative accuracy of the estimation of water concentration has increased to 94% compared to the HITRAN fitting. Meanwhile the accuracy of the ammonia and methane concentrations has decreased to 62% and 48%, respectively. This decrease in accuracy is due to the relativity low concentration of ammonia and methane. For methane it is mainly because of the mismatch between the wavelength axis of the test spectra and the training spectra. Comparing
Figure 5b with the training data in
Figure 2c we find that the experimental PAS spectra is shifted approximately 2 nm to the left in the 3.2 to 3.35
m range. This can be seen from
Figure 5b and is the reason that the PLS fitting for methane becomes very inaccurate and underestimated the presents of methane. Note that the experimental conditions are very similar for the experimental PAS spectra shown in
Figure 3b, where the wavelength deviation is less than 0.2 nm. We apply the same three gas trained PLS method to the data in
Figure 3b and find concentrations of 7196 ppm/V, -4 ppm/V and 30 ppm/V for water, ammonia and water, respectively. Thus, we conclude that the main reason for the inaccuracies shown in
Figure 5b is due to the deviation of wavelength axis compared to the training/HITRAN spectra and that in order for the PLS to estimate with high accuracy the deviation in wavelength should be less than 1 nm. In
Figure 5c,d the deviation is less than 0.5 nm and we continue with the investigation of how very asymmetric mixing ratio affect the accuracy of the PLS method. In this case the concentrations of ammonia and methane becomes relative low compared to the water content. The water spectra will therefore dominate the trained PLS model due to the high level of spectral overlap with ammonia and methane. To test this behavior and make an estimation of the PLS methods lower accuracy (sensitivity) limit we increase the amount of water and ammonia. Thus diluting the content of methane at the same time. This is depicted in
Figure 5c where we find that the PLS method is making a completely wrong estimation of the concentration by estimating it to be negative (-5 ppm/V). Due to this apparent wrong estimation of the methane concentration the PLS method predicts a 3% higher concentration of water relative to the HITRAN fitting. This fitting behavior is in general very different from the fitting behavior that we have seen in the previous cases in
Figure 3 and
Figure 4, where the HITRAN fitting always estimated higher concentrations. It suggests that for high concentrations of water and for relatively low concentrations of ammonia and methane the PLS algorithm are using the ammonia and methane coefficients as a free parameter to make a better fit of the water spectra. This is clearly seen from the fact that the relative precision between the PLS method and HITRAN fitting becomes higher in this case. Therefore to estimate the accuracy and sensitivity we dilute the concentration of water and ammonia by purging with methane as shown in
Figure 5d. It can be seen than the PLS method now gives only positive coefficients for the estimation of the concentrations. However, the estimations of ammonia and methane are still underestimated relative to the HITRAN fitting while the water is overestimated with 2%. If we apply the enhancement and correction factors found in
Figure 4 for the calibration of absolute concentrations of methane, we find a methane concentration of 3 (±2) ppm/V and 5(±2) ppm/V for the PLS method and HITRAN fitting, respectively. Thus, the estimations of the contractions with the two methods are in good agreement within the measurements uncertainties. By applying the same water correction factor from
Figure 4b to the estimated ammonia concentration we find that the absolute concentration should be approximately 50(±10) ppm/V and 67(±10) ppm/V for the PLS method and HITRAN fitting, respectively. We conclude from these measurement and tests that the absolute sensitivity of the three gas training PLS is approximately 300(±50) ppm/V, 50(±5) ppm/V and 5(±2) ppm/V for water, ammonia and methane, respectively.