In addition to wavelength interrogation mode, the sensor can also operate in gate voltage interrogation mode: i.e. detecting the change in chemical potential of graphene by scanning the gate voltage provided by a high resolution digital-to-analog converter (DAC). The reflectance versus gate voltage SPR curve can be obtained through calculation as follows: according to the wavelength of monochromatic incident light beam λ’, the corresponding resonance angle θ(λ’) (see
Figure 5) is selected as the fixed incident angle, the chemical potential of graphene is adjusted varying around the corresponding PR point μ
cPR(λ’) (see
Table 1), the refractive index of graphene within the range of chemical potential change can be calculated by using Eq. (1) and (2) under the conditions of λ=λ’, T=298 K, Γ=1e12 Hz, and the gate voltage corresponding to the varying μ
c can also be calculated by using Eq. (4) and (5). If the initial chemical potential of graphene is 0, that is, the undoped state, then the minimum reflectance point of the SPR curve will occur at the corresponding gate voltage required for the graphene chemical potential to reach the PR point (see
Table 1), which may be referred to as the resonance gate voltage. When the initial chemical potential changes from 0 due to the attachment of biomolecules, the minimum reflectance point of the SPR curve is shifted because part of the contribution of the gate voltage applied to bring the chemical potential to the PR point is replaced by the contribution of the attached biomolecules. So the shift of the gate voltage at minimum reflectance point (i.e. resonance gate voltage) can be used for indicating the change of the initial chemical potential, and consequently indicating the attachment of target biomolecules. For the monochromatic incident light with wavelengths of 1100, 1200, 1310, 1550, 1700 and 1900 nm, using fixed incident angles of 71.7°, 80.3°, 84.9°, 86.8°, 86.2° and 85.1° respectively, and adopting the optimal materials of prism for different wavelengths, i.e. LF for 1100 nm, CaF
2 for 1200 nm, MgF
2 for 1310, 1550, 1700 and 1900 nm respectively, the evolution of reflectance versus gate voltage SPR curves with the change of the initial chemical potential in graphene film are depicted in
Figure 10. Here we assume that both the added gate voltage and the attached biomolecules exert same doping effect on graphene, i.e. both N-doping or both P-doping. It can be seen that with the increase of the initial chemical potential (absolute value), the resonance gate voltages all shift to smaller values. When using 1100 nm as the incident light wavelength, care should be taken not to make the scanning gate voltage exceed 57.5 V, otherwise the dielectric layer would be broken down, as shown in
Figure 10a. We can calculate the CP sensitivities in gate voltage interrogation mode, S’
CP, defined as the ratio of the change in resonance gate voltage to the change in initial chemical potential of the graphene film ΔV
gres/ΔCP. At the incident wavelengths of 1100, 1200, 1310, 1550, 1700 and 1900 nm, the values are 156.9822, 143.6147, 131.0779, 111.0351, 101.3415 and 90.6038 mV/meV respectively. The FWHMs can also be calculated, and the values are 1158, 1394, 1475, 1790, 1522 and 1474 mV respectively at the incident wavelengths of 1100, 1200, 1310, 1550, 1700 and 1900 nm. According to Eq. (5), a same change of chemical potential in a higher chemical potential value region leads to a larger change of the charge carrier density (thus leading to a larger change in the gate voltage that to be applied). As the μ
cPR decreases with the increase of wavelength (see
Table 1), CP sensitivity (S’
CP) decreases with the increase of incident light wavelength, as shown in
Figure 11. For the chemical potential shift of >10 meV caused by the hybridization of 0.01 nM complimentary DNAs to the probe DNAs pre-immobilized on the graphene [
62], the resonance gate voltage would shift >1569.822, >1436.147, >1310.779, >1110.351, >1013.415 and >906.038 mV respectively at the incident wavelength of 1100, 1200, 1310, 1550, 1700 and 1900 nm. The minimum detection limit of the sensor in bio sensing under gate voltage interrogation pattern is related to the resolution of DAC. The output voltage of DAC is in fact discrete with a minimum step equal to δV
DAC=V
REF/(2
M-1), where M is the bit number of a DAC and V
REF is the reference voltage (full scale output) [
40]. Since the maximum gate voltage should not exceed 57.5 V, the reference voltage is set to be V
REF=57 V. For a 24-bit DAC, δV
DAC=3.4 μV. Suppose the proposed near-IR GSPR sensor is used in DNA sensing described in Ref. [
62], it can reach surprising theoretical LOD of <21.66, <23.67, <25.94, <30.62, <33.55 and <37.53 aM in gate voltage interrogation mode respectively at the incident wavelength of 1100, 1200, 1310, 1550, 1700 and 1900 nm. If a DAC with a larger bit number M is adopted, it is conceivable that the LOD will decrease further.