Method validation
Subclause 7.2 of ISO 17025:2017 suggests all necessary parameters need to be checked for method validation.9 However, in this paper, only the five parameters defined below were majorly focused on due to the limitation of our laboratory’s resources. In addition, the definitions of them in this paper are a bit different than those in ISO 5725-1:1994.10 Hereby, to avoid unnecessary ambiguity, we will re-clarify their definition and describe the method of how to demonstrate them in this work before the validation discussion.
Definitions
Accuracy11 or trueness10, measures the nearness to the truth.11 This work demonstrated it by comparing results from two or more different analytical methods and expected to agree within their expected precision.
Precision shows how well replicate measurements agree with one another.11 In this case, it was evaluated by calculating the repeatability and the Intermediate precision11 (same laboratory, same instrument, same method but different operators and different time) of the method. Since it was inapplicable to perform reproducibility (different laboratory, different instrument, and different operator) evaluation, the intermediate precision was treated as Reproducibility and calculated using the formulas in subclause 7.4 of ISO 5725-2:1994.12
Linearity evaluates how well a calibration curve follows a straight line, which demonstrates the proportional relationship between the signal and the amount of analyte.11
Lowest limit of detection (LoD) is the smallest quantity of analyte that is significantly different from the blank but not enough for accurate measurement.11 Signal-to-noise ratio S/N = 3 for LoD.
Lowest limit of quantification (LoQ) is the smallest amount that can be measured with reasonable accuracy.11 Signal-to-noise ratio S/N = 10 for LoQ.
The calibration curves in
Table 2 were derived by using the method of least square in which the sum of squares of the vertical deviations between the data points and the line is minimized.
Calibration curve: , where y is the XRF signal, x is the concentration of analytes, m is the slope of the linear line.
Vertical deviation for the point () =
The standard deviation of vertical deviation = ,where n is the number of data points
Thereby, the signal of detection limit was estimated to be equal ,11 hence the LoD = , and LoQ =
Validation results
This method was extrapolated from the assumption that the matrix between the standard solution (5% HCl) and sample (DI water with a small amount of methanol and PVP) resembles. To test the hypothesis, a t-test for comparing two different sets of replicate measurements was performed. As can be seen in
Table 2, the calculated t-value were all below the theoretical t-value, indicating that the sets of data from 5% HCl matrix and sample matrix agree each to other. These results prove that the two backgrounds of analyses are not significantly different.
In
Table 1, the RSDs of the quantitative results are below 5,3%, which is the predicted standard deviation for repeatability at 100 ppm of mass fraction (see AOAC appendix F: SMPR Guidline),
13 demonstrating that the method has extremely high repeatability. For evaluating reproducibility, the reproducibility variance
12 of PdRh results from two measurements six months away from each other.
Table 3 reported the relative intermediate precision variance (
) of 2,35% that is much lower than the predicted standard deviation, 8%, for reproducibility at 10
-4 of mass fraction,
13 indicating the method to be highly reproducible. Overall, this quantitative method possessed relatively high precision.
Moreover, the square of correlation coefficients R
2 of all five calibration curves are all higher than 0.995,
11 which indicates good linearity (
Table 4). In addition, Pt and Ir had slopes of 3,9548 and 2,6228, which are much higher than others, showing that these heavy elements have a higher sensitivity than of the lighter ones.
In
Table 4, The LoDs of all five target elements are above 1 mg/kg which was rational since the published LoD from the instrument manufacturer is 1ppm (roughly 1mg/kg).
14 Besides, the LoQ of all five elements is suitable for quantifying common nanoparticle solution synthesized in research laboratories. Thus, the XRF quantitative results in
Table 1 are reliable.
To evaluate the accuracy, XRF quantitative results were converted to the molar ratio form. Hereafter, the molar ratio results from SEM-EDS and the XRF calibration method were taken under a paired t-Test for comparing individual differences by assuming that SEM-EDS results are more reliable since the uncertainty of SEM-EDS results were much more narrow than of XRF calibration method. In the matter of fact, SEM-EDS is a very straight-forward analysis without any redundant sample preparations that might interfere to the final results.
In
Table 5, The calculated t was roughly 1,523, which is lower than the theoretical t value 2,571 at 95% confidence. Thus, there is more than 5% chance, the two sets of results lie within experimental error. Or in another word, the results from these two methods agreed to each other. This finding demonstrated that it is possible to quantify bimetallic alloy solution by just quantifying one of two elemental components in the alloy. Therefore, this interchangeability with SEM-EDS allows the XRF calibration method to be compatible with high-throughput (HTP) experiments, which generate a massive amount of data in a short time.
Last but not least, in fact, both the XRF calibration method required a very small amount of sample without conducting any sample digestion. Furthermore, by the intrinsic ability to analyze simultaneously multi elements of XRF and SEM-EDS, and the application of multi-elements standards, analysis time can be significantly shortened. These lead to the other several great advantages of this method, which are non-destructive, economical, and fast. The advantages vastly enhance the method’s compatibility with HTP experiments.