The three setups covered in this work are depicted in
Figure 3. The single-pixel setup (Setup 1,
Figure 3(a)) features one transmitter and one receiver applying the frequency-modulated continuous wave (FMCW) principle to emit and detect terahertz radiation sweeping in a range from 230 to 320 GHz. They are equipped with Pickett-Potter horn antennas emitting a beam, which is collimated by a PTFE lens (
and focused to the center of rotation by a second
lens. The lens configuration shapes the beam approximately into a Gaussian beam with a beam waist of
. The receiver unit, which features the same lens setup, mixes the detected radiation with a local oscillator signal, coherently acquiring the signal amplitude and phase. This allows a determination of a sample’s full refractive index from the measured amplitude reduction and time delay of the signal. The single-pixel setup is described in more detail in [
14,
15]. In [
15] we presented an algorithm to include
a priori information and off-axis measurements in the tomographic reconstruction process. The latter requires time-costly scans of the whole detector plane, which have to be acquired by moving the single-pixel detector along the whole range of motion for every transmitter position and angle. To reduce this acquisition time, we developed Setups 2 and 3.
In Setup 2, displayed in
Figure 3(b), the receiver unit is replaced with an eight-channel detector array, operating at the same frequency range as System 1 (230 to 230 GHz) [
18]. A photograph of the eight-channel array is presented in
Figure 2. Analogous to the single-pixel receiver in System 1, each channel features a Picket-Potter horn antenna, which channels the received terahertz signal to a subharmonic mixer fed secondly with a local oscillator signal stemming from the voltage-controlled oscillator in the transmitter. One local oscillator signal is sufficient to allow the simultaneous homodyne signal acquisition of eight detection channels. With this setup, we can potentially achieve an increase in measurement velocity by a factor of eight when acquiring scans of the whole detector plane. However, it still requires a stepped movement of the transmitter. To render this movement of the transmitter unnecessary and increase the imaging speed even further, we introduce a telecentric f-
lens combined with a rotating scanning mirror, building up Setup 3.
Figure 2.
The detector array employed in Setup 2 and 3 features eight independent channels with one antenna and one homodyne mixer each. It is fed with a local oscillator signal and performs coherent detection of eight pixels simultaneously. The width of each antenna is 8 mm, so that the whole array spans 64 mm in total.
Figure 3.
Tomographic setups compared in this work: (a) shows the single-pixel setup with single transmitter and receiver unit moveable independently from each other. Two lenses in front of the transmitter collimate and focus the beam in the center of the sample. The receiver features the same lens setup to improve detection. (b) The single receiver pixel is replaced by an 8-channel array increasing the number of acquired pixels at a time.To increase the resolution in the detection plane the array is moved in a step-wise fashion, and so is the transmitter as well. In (c) the necessity of moving the transmitter is circumvent with a combination of a scanning mirror and an f- lens replacing the focussing lens of the transmitter optics. The collimated beam is guided by the mirror onto the f- lens from different angles, leading to a parallel shift of the focused beam.
An important difference between Setup 1 and the following two setups is the lack of focusing optics in front of the detector array, which is difficult to implement due to limited physical space between the densely positioned array antennas. As a result, we expect less directivity of the detector pixels and ultimately a lowered resolution in the reconstructed images as a price to pay for the increased measurement speed.
2.1. Design of the f-θ optics
f-θ lenses are commonly used in many applications involving the displacement of a laser beam for scanning, engraving [
20], optical coherence tomography (OCT) [
21], or material processing applications. Unlike the common spherical lens, which has a curved focal plane, an f-θ lens provides a flat focal plane with low field curvature, a large field of view, and homogenous beam characteristics in the image plane throughout the entire scan field.
These objective lenses are increasingly applied in microwave and terahertz applications for 3D imaging [
22], food safety inspection [
23], non-destructive testing [
24], etc. Thanks to the non-dispersive characteristics of the available lens fabrication material in terahertz and sub-terahertz, f-θ lenses are incorporated with various transmitters and receivers operating in this frequency region, such as FMCW radars with varying bandwidths [
22,
24,
25] and TDS systems up to 1.25 THz [
26].
In this work, the f-θ lens is designed and optimized for an effective focal length of
, a diameter of
, and a scan line of
, as shown in
Figure 4. It is fabricated out of high-density polyethylene (HDPE) using computer numerical control (CNC) milling. The scanning mirror moves
, deflecting the beam onto the f-θ lens with
. The displacement of the probing beam from the optical axis is proportional to the incident angle
:
Two objectives were the main criteria for the design of the lens:
Creating a flat focal plane with telecentricity in the image plane, which guarantees a normal incidence of the focused beam onto the sample.
Ensuring that the focus at each measurement point is as small as possible, ideally only limited by diffraction, and for the beam shape to be consistent throughout the mirror angles θ.
The lens design includes aspherical surfaces to eliminate geometrical aberrations. Before entering the f-θ lens, the beam is collimated using a lens. The beam size after collimation has been measured to be approximately . We assess the quasi-optical performance of the objective lens by validating the Gaussian beam propagation. To do so, we simulate the optical behavior of the f-θ lens with the optical design software Zemax, modeling the horn antenna and the collimating lens as ideal components.
In
Figure 5, we compare raster scans of Gaussian beams created by the f-θ lens under incidence from different angles
with a scan of the reference beam prduced by the standard PTFE focusing lens employed in Setup 1 and 2. The raster scan images in
Figure 5 are acquired by moving the detector array stepwise along the beam axis and perpendicular to it. As expected, the beam focus lies in the origin of the diagram, the abscissa represents the direction of the beam propagation. The beam waist is defined by a decay of intensity to
(indicated by the green lines in the figure) is consistent throughout the range of
. With a values of around
, the measured focus beam widths of the Gaussian beams leaving the f-θ lens are slightly wider than the beam width of the reference beam of
(
Figure 5(a)). As an example, we show three images of the beams for mirror angles between
and
in
Figure 5(b)-(d). The difference in beam width between the systems will presumably result in a lowered resolution in the reconstructed images. Especially for very excentric mirror positions, the beams tend to widen slightly before the focus
in
Figure 5(b) and (d). Nevertheless, for
all beams show an intensity distribution, that is consistent with the Gaussian beam approximation with
. This indicates, that the f-θ lens enables a scan of the imaging scene with beams focused in the imaging plane, that can be successfully described by the Gaussian beam model.
2.2 Image reconstruction process
The image reconstruction employed in this work has been discussed at length in [
14], and the incorporation procedure of
a priori information is covered comprehensively in [
15]. Here, we will therefore only give a brief introduction to the aforementioned concepts. The reconstruction algorithm is based on the Conjugate Gradient Least Square algorithm (CGLS) [
19]. It is a fast, versatile technique to approximate iteratively the solution to the inverse problem
In this case
represents the unknown image vector and
stands for the measured data acquired in the measurement process, also referred to as a sinogram. The matrix
models the path on which the probing beam of terahertz radiation travels through the imaging scene. When on this path
, the beam interacts with an object, which has a refractive index of
differing from the refractive index of air
. The time of flight (TOF) is directly proportional to the real refractive index
whereas the relative intensity loss
is according to Lambert-Beer’s law:
Inserting the definition of the absorption coefficient
, we can express the intensity loss in its logarithmic representation as:
Discretizing the imaging scene
as a square area of
pixels, (4) and (6) become
Here represent the length of the path on which the beam crosses pixel . and are assumed to be constant within one pixel. The column vectors of the matrix are populated with the values, determining which pixels contribute to which beam to which extent. This way, by inserting the measured TOF values into the sinogram vector in (2), we can solve for the real refractive index distribution of the imaging scene, and for the image showing the absorption coefficient, respectively.
The flexibility of the CGLS algorithm comes with the possibility to freely design the Matrix
. As it is explained thoroughly in [
15], this opens up the opportunity to incorporate a priori information into the tomographic reconstruction process, allowing more precise modeling of the measurement process. While the above model of the interaction between the beam and the sample is based on the assumption that the probing beams had a vanishing diameter, it is in fact a Gaussian beam with a diameter relevant to the feature size of the objects (see
Section 2.1). Ignoring this leads to significant distortions and reconstruction errors in the image [
15]. We were able to incorporate the non-vanishing beam size into the matrix
by considering one beam as a combination of one-dimensional infinetly thin rays with different starting positions and directions (see
Figure 6). This model describes the probing beam more accurately but preserves the linear relation between the sinogram values and the image pixels in (7) and (8).
The second set of a priori information we consider in the reconstruction process is the shape of the sample and its refractive index. In the field of NDT, this is a relevant scenario, since the desired shape and material of a sample are often known a priori. Since the wavelength of the terahertz radiation in use is in the same order of magnitude as the features to be observed, optical effects occurring at the interfaces between the sample and the surrounding air can influence the beam propagation severely. These optical effects include refraction and reflection, which alter the beam propagation directions according to Snell’s law. Considering the outer boundaries of a sample as well as its refractive index, one can apply ray tracing to predict the ray paths on which the radiation travels through the imaging scene (see
Figure 6).
Figure 6 displays, that for certain sample geometries and angles, the probing radiation can be deviated far from its originally straight paths. To be able, nevertheless, to consider the deviated radiation in the reconstruction process, we conduct off-axis measurements, i.e., a scan of the whole detector plane for deviated radiation. From the ray-tracing simulations, we determine the position of the highest intensity and include the values measured at this position in the reconstruction as described in [
15]. While this procedure increases the quality of the reconstructed images, it renders the acquisition very time-consuming, when working with Setup 1, because one has to scan the whole detector plane stepwise or in a continuous movement for every transmitter position (and rotation angle). Hence, we developed Setup 2 and finally Setup 3 to reduce the acquisition time. The former facilitates the simultaneous acquisition of data at eight different positions in the detection plane. The latter allows a continuous acquisition at eight different positions in the detector plane, while the rotation of the mirror leads to scanning beams moving through the emitter plane.