1. Introduction
Photodynamic therapy (PDT) and photoimmunotherapy (PIT) are treatment methods that involve the use of a tumor-selective photosensitive substance (PS) and laser light with a wavelength that matches the absorption of the PS to induce a photochemical reaction and eliminate cancer cells [
1,
2]. In PDT, the PS and laser light are used to target and destroy tumor cells [
3], while in PIT, the PS in combination with tumor specific antibody is used to target cancer cells and the reaction in PIT leads to activate the immune system [
4]. Although both PDT and PIT are effective treatment options for cancer, they have different mechanisms of action and may be more appropriate for specific types of cancer. In clinical practice, PDT using talaporphine sodium has been approved and implemented for the treatment of locally recurrent esophageal cancer, early stage lung cancer, and primary malignant brain tumors following chemoradiotherapy or radiation therapy [
5]. Currently, a phase III clinical trial is underway in patients with head and neck cancer utilizing cetuximab-IR700 (Cet-IR700) to target the epidermal growth factor receptor (EGFR) [
6,
7]. This drug has been approved in Japan for the treatment of unresectable locally advanced or locally recurrent head and neck cancer based on the outcomes of a phase I trial [
8]. Investigator-initiated clinical trials are currently being conducted to evaluate the use of PIT for localized esophageal cancer and unresectable advanced or recurrent gastric cancer resistant to conventional therapies [
9].
The successful delivery of PDT and PIT relies on the ability to activate the PS with sufficient light energy while selectively illuminating the tumor. However, excessive light energy can cause damage to normal tissues, resulting in serious adverse effects [
10,
11]. Conversely, insufficient light energy may reduce the treatment effectiveness. Therefore, it is critical to develop methods that accurately detect the extent of tumors and to determine the necessary and sufficient light irradiation dose to exert therapeutic effects. Previous research by Sato et al. reported that the fluorescence of Cet-IR700 can be observed by NIR light exposure [
12,
13]. Takashima et al. used a fluorescence imaging system (LIGHTVISION, SHIMADZU, Japan) to visualize real-time drug accumulation in tumors for light-sensitive substances, and they demonstrated a correlation between the decay rate of the fluorescence intensity and the anti-tumor effect [
14]. However, current endoscopes are unable to employ the configuration required for fluorescence observation when treatment is performed especially in digestive tract.
The proposed device, which integrates a narrow-diameter camera and laser diffuser, represents a novel technique for fluorescence imaging introduced through the forceps port. This device has the potential to facilitate the administration of appropriate doses of fluorescence intensity while performing laser irradiation under general endoscopy, thereby optimizing PDT and PIT treatment. This study focused on the development of the proposed device and investigated the feasibility of visualizing PIT tumor accumulation and monitoring fluorescence intensity decay in vivo.
2. Materials and Methods
2.1. Development of Fluorescence Imaging Camera with Laser Diffuser
This study involved the design and construction of a structure, as depicted in
Figure 1, which was capable of being inserted into an endoscope with a φ3.2-mm forceps port. The structure was designed to emit laser light and acquire fluorescent images simultaneously. The sensor component of the structure was a 1 mm × 1 mm near-infrared camera (NIRBP, Opta Sensor, Germany) equipped with a bandpass coating (813-878 nm transmission, OD4). The camera specifications were F2.7, FOV90º, 320 × 320 pixel, 10-bit depth, and a 10-ms exposure time. A 690-nm notch filter (672-698-nm block, OD6, Edmund, USA) was attached to the tip of the camera to block high-intensity laser light. To prevent noise, dark images were prestored, and the acquired images were subtracted from the dark images.
A frontal diffuser (FD1; Medlight S.A., Switzerland) was used as the laser diffuser. The tip of the diffuser was covered with a Teflon tube (OD: 2 mm, ID: 1.5 mm), from which approximately 5 mm had been peeled from the top. The joint between the laser diffuser and the NIR camera was fabricated using a three-dimensional (3D) printer (XFAB 2000, DWS, Italy) and a Therma284. The 2-m-long flat cable for the camera and laser diffuser was connected using a heat-shrinkable tube (NR0502-003, Flonchemical Co. Ltd., Japan).
2.2. Real-Time Fluorescence Imaging Using Test Sample by Developed System
To verify the fluorescence imaging functionality, negative and positive control liquids were analyzed using the developed system. Water was used as the negative control, whereas IRDye 700DX NHS Ester (IR700; Lincoln, NE, USA) diluted 50 times with PBS served as the positive control. The excitation light source was a 690-nm laser (150 mW/cm2) from a laser irradiation device (MLL-III-690, Changchun New Industries Optoelectronics Technology Co., Ltd., China) used in PIT therapy. The probe head was positioned approximately 34 mm from the observation target. The acquired fluorescence image was black and white; thus, it was assigned a green color in the color image.
2.3. Animal Model
To obtain tumor-bearing mice, the A431 cell line was obtained from the American Type Culture Collection (Manassas, VA, USA). The cell line was cultured in Dulbecco’s modified Eagle’s medium (DMEM) (FUJIFILM Wako Pure Chemical Corporation, Osaka, Japan) supplemented with 10% fetal bovine serum (FBS, Thermo Fisher Scientific, Waltham, MA, USA) and 1% penicillin-streptomycin-amphotericin B suspension (FUJIFILM Wako Pure Chemical Corporation, Osaka, Japan). The cells were maintained at 37 °C and 5% CO
2 in a humidified incubator. Two female BALB/c-nu nude mice at 6 weeks of age (Charles River Laboratories Japan, Inc., Japan) were implanted with 3.5 × 10
6 A431 cells in the right waist. The tumor volume was calculated using the following equation: TV = (L × W
2)/2, where L and W are the length and width of the tumor under the skin, respectively [
15]. All the procedures and protocols were approved by the Animal Care and Use Committee of the National Cancer Center (K21-010).
2.4. In Vivo Fluorescence Imaging Using Developed System
Cet-IR700 was used as an IR700-conjugated antibody for administration in tumor-bearing mice. Specifically, two mice with A431 tumors of approximately 100 mm3 were administered with 100 μg of Cet-IR700, followed by laser irradiation 24 h later for 330 s. During this process, fluorescent images of the tumor region were captured. Pixel values were measured every 2 s to create a graph correlating light intensity with treatment time. After treatment, a halogen lamp (JCR15V150WS and LA-150UE, Hayashi-Repic Co., Ltd., Japan) was exposed, and image acquisition was performed to merge the fluorescence image with the overall image of the mouse. To enhance the hue, the intensity of the fluorescence image was doubled, assigned to green, and integrated into the image irradiated with a halogen lamp.
Figure 2.
(a) Fluorescence imaging acquisition setup using mouse, (b) Cet-IR700-dosed tumor-bearing mice, and (c) Laser diffuser irradiation method.
Figure 2.
(a) Fluorescence imaging acquisition setup using mouse, (b) Cet-IR700-dosed tumor-bearing mice, and (c) Laser diffuser irradiation method.
3. Results
We successfully developed a device that integrates a narrow camera and a laser diffuser, as depicted in
Figure 3a. This device has the ability to reduce the excitation wavelength of the laser and has a diameter of less than 3 mm, allowing it to be inserted into a 3.2-mm-diameter channel using the forceps port on an endoscope, as illustrated in
Figure 3b.
This device was used to image phantoms containing water and the IR700 solution, as shown in
Figure 4a,d. As shown in
Figure 4b,c, no fluorescence was detected in the water sample, whereas fluorescence was observed in the IR700 tube when exposed to the laser, as shown in
Figure 4e,f. This confirmed that the laser wavelength was blocked by the filter, and only the fluorescence wavelength was captured.
Fluorescence was detected in two mice (A431) administered cetuximab-IR700 and exposed to a 690-nm laser (150 mW/cm
2). As depicted on the left sides of
Figure 5a,b, fluorescence was evident immediately after irradiation in both cases. When merged on an image illuminated by a halogen lamp, as shown on the right side of
Figure 5a,b, the fluorescence of the tumor region was visible. After 330 s of irradiation, as shown in
Figure 5c,d, the fluorescence diminished and disappeared. To quantify the fluorescence intensity over time, the tumor area was designated as the region of interest, as shown in
Figure 6a,b. The number of pixels was 10105 and 4557 px, respectively. The mean and standard deviations are plotted in
Figure 6c,d, respectively, and it was observed that in both mice, fluorescence rapidly declined immediately after irradiation; the decay of the fluorescence intensity reached equilibrium after approximately 300 and 250 s, respectively.
4. Discussion
Takashima et al. conducted a study using nine mice in a fluorescence imaging experiment and found that the fluorescence decayed rapidly after irradiation and reached equilibrium between 150 and 330 s, maximizing the anti-tumor effect [
14]. As the present results exhibited similar behavior, the obtained fluorescence images were considered reliable. Previous research has demonstrated a relationship between the required laser dose and fluorescence intensity, emphasizing the significance of visualizing the therapeutic effects. Ongoing observational studies on head and neck surgery aim to confirm fluorescence attenuation in humans (UMIN000052240). Moreover, interest extends to endoscopic observation, with potential applications in the medical field. In this study, the feasibility of observing drug accumulation in tumors under endoscopy and identifying a sufficient irradiation time for PS was demonstrated using the developed device. This device has the potential for application not only for PIT but also for the fluorescence imaging of PDT and indocyanine green (ICG) under endoscopy.
To introduce the device into clinical practice, improvements in visibility and usability include a wider viewing angle, a new optical filter design, and a structure that allows the tip of the device to be curved. Although adopting a wide-angle lens may result in a reduction in resolution, our results indicate that tumor volumes of 100 mm3 or more can still be effectively identified as fluorescent regions, even with a resolution decrease of up to two or three times. The typical field of view of an endoscope is approximately 140°, whereas the camera used in this study had a field of view of 90°. Thus, by integrating a wide-angle objective lens, the observations can be compared with an endoscopic view.
With respect to the optical filter section, only cameras coated with a bandpass filter of OD4 are commercially available. Consequently, a notch filter of OD6 was added to this development. However, notch filters may be unnecessary when using an order-made image sensor coated with a higher-optical-density filter. Regarding the need for a curved structure, it is difficult for the frontal diffuser to irradiate in the lateral direction with only endoscopic curvature; however, it is difficult to install a fluorescence imaging sensor with a cylindrical diffuser. Therefore, the use of a curved tip device such as the SpyScope™ biliary speculum is expected to enable lateral irradiation [
16]. This structure could be realized by making the laser diffuser thinner and softer. One of the potential limitations of the proposed imaging method is the difficulty in registering the observed fluorescence onto the endoscopic image. This issue may be resolved by image registration processing, such as an affine transform, before endoscopy.
The proposed configuration can be deployed as a single-use device because a disposable camera is mounted on a laser diffuser, which is disposable in clinical use. Moreover, because there is no need to add a new filter for endoscopes, and the PS effect can be visualized regardless of the manufacturer of the endoscope, the system is considered economically viable, and is expected to be applied clinically in the near future.
5. Conclusions
In this study, we successfully developed a laser irradiation device with a built-in near-infrared fluorescence imaging sensor that can be introduced into the forceps opening of an endoscope. The imaging sensor transmitted only the fluorescence wavelength, and it was confirmed that the fluorescence of Cet-IR700 that accumulated in the tumor could be visualized. The decay of the fluorescence intensity by laser irradiation was measured using the developed device, and it was confirmed that the intensity decreased with irradiation time and reached equilibrium at approximately 330 s. This behavior is similar to that observed in a previous study that used a fluorescent camera (LIGHTVISION) to visualize the fluorescence of PIT in mice. Thus, we show the visualization of the decay of PIT fluorescence under an endoscope in a clinical setting.
Author Contributions
T.T. and T.Y. designed and conducted the research. H.T. performed the in vivo experiments. T.T. performed image data analysis. T.T. and T.Y. interpreted the data. T.T. wrote the paper. All the authors reviewed the manuscript.
Funding
This research was funded by the National Cancer Center Research and Development Fund (grant number 2023-A-9).
Institutional Review Board Statement
All procedures and protocols were approved by the Animal Care and Use Committee of the National Cancer Center (K21-010).
Informed Consent Statement
Not applicable.
Data Availability Statement
The data presented in this study are available upon request from the corresponding author.
Conflicts of Interest
The authors declare no conflicts of interest.
References
- Dolmans, D.E.J.G.J.; Fukumura, D.; Jain, R.K. Photodynamic therapy for cancer. Nature Reviews Cancer 2003, 3, 380–387. [Google Scholar] [CrossRef]
- Mitsunaga, M.; Ogawa, M.; Kosaka, N.; Rosenblum, L.T.; Choyke, P.L.; Kobayashi, H. Cancer cell–selective in vivo near infrared photoimmunotherapy targeting specific membrane molecules. Nature Medicine 2011, 17, 1685–1691. [Google Scholar] [CrossRef] [PubMed]
- Litle, V.R.; Luketich, J.D.; Christie, N.A.; Buenaventura, P.O.; Alvelo-Rivera, M.; McCaughan, J.S.; Nguyen, N.T.; Fernando, H.C. Photodynamic therapy as palliation for esophageal cancer: Experience in 215 patients. The Annals of Thoracic Surgery 2003, 76, 1687–1693. [Google Scholar] [CrossRef] [PubMed]
- Maruoka, Y.; Wakiyama, H.; Choyke, P.L.; Kobayashi, H. , Near infrared photoimmunotherapy for cancers: A translational perspective. EBioMedicine 2021, 70, 103501. [Google Scholar] [CrossRef] [PubMed]
- Yano, T.; Minamide, T.; Takashima, K.; Nakajo, K.; Kadota, T.; Yoda, Y. Clinical Practice of Photodynamic Therapy Using Talaporfin Sodium for Esophageal Cancer. Journal of Clinical Medicine 2021, 10, 2785. [Google Scholar] [CrossRef] [PubMed]
- Biel, M.A.; Gillenwater, A.M.; Cognetti, D.M.; Johnson, J.M.; Argiris, A.; Tahara, M. A global phase III multicenter, randomized, double-arm, open label trial of ASP-1929 photoimmunotherapy versus physician’s choice standard of care for the treatment of patients with locoregional, recurrent head and neck squamous cell carcinoma (rHNSCC). Journal of Clinical Oncology 2019, 37 (Suppl. 15), TPS6094–TPS6094. [Google Scholar] [CrossRef]
- Miyazaki, N.L.; Furusawa, A.; Choyke, P.L.; Kobayashi, H. Review of RM-1929 Near-Infrared Photoimmunotherapy Clinical Efficacy for Unresectable and/or Recurrent Head and Neck Squamous Cell Carcinoma. Cancers 2023, 15, 5117. [Google Scholar] [CrossRef]
- Tahara, M.; Okano, S.; Enokida, T.; Ueda, Y.; Fujisawa, T.; Shinozaki, T.; Tomioka, T.; Okano, W.; Biel, M.A.; Ishida, K.; Hayashi, R. A phase I, single-center, open-label study of RM-1929 photoimmunotherapy in Japanese patients with recurrent head and neck squamous cell carcinoma. International Journal of Clinical Oncology 2021, 26, 1812–1821. [Google Scholar] [CrossRef] [PubMed]
- Kadota, T.; Kotani, D.; Yoda, Y.; Fukutani, M.; Wakabayashi, M.; Nomura, S.; Fuse, N.; Sato, A.; Yano, T.; Shitara, K. A phase Ib study of near infrared photoimmunotherapy (NIR-PIT) using ASP-1929 in combination with nivolumab and for patients with advanced gastric or esophageal cancer (GE-PIT study, EPOC1901). Journal of Clinical Oncology 2020, 38 (Suppl. 4), TPS457–TPS457. [Google Scholar] [CrossRef]
- Mitsunaga, M.; Nakajima, T.; Sano, K.; Kramer-Marek, G.; Choyke, P.L.; Kobayashi, H. Immediate in vivo target-specific cancer cell death after near infrared photoimmunotherapy. BMC Cancer 2012, 12, 345. [Google Scholar] [CrossRef] [PubMed]
- Okuyama, S.; Nagaya, T.; Sato, K.; Ogata, F.; Maruoka, Y.; Choyke, P.L.; Kobayashi, H. Interstitial near-infrared photoimmunotherapy: Effective treatment areas and light doses needed for use with fiber optic diffusers. Oncotarget 2018, 9, 11159–11169. [Google Scholar] [CrossRef]
- Sato, K.; Ando, K.; Okuyama, S.; Moriguchi, S.; Ogura, T.; Totoki, S.; Hanaoka, H.; Nagaya, T.; Kokawa, R.; Takakura, H.; Nishimura, M.; Hasegawa, Y.; Choyke, P.L.; Ogawa, M.; Kobayashi, H. Photoinduced Ligand Release from a Silicon Phthalocyanine Dye Conjugated with Monoclonal Antibodies: A Mechanism of Cancer Cell Cytotoxicity after Near-Infrared Photoimmunotherapy. ACS Central Science 2018, 4, 1559–1569. [Google Scholar] [CrossRef] [PubMed]
- Inagaki, F.F.; Fujimura, D.; Furusawa, A.; Okada, R.; Wakiyama, H.; Kato, T.; Choyke, P.L.; Kobayashi, H. Diagnostic imaging in near-infrared photoimmunotherapy using a commercially available camera for indocyanine green. Cancer Science 2021, 112, 1326–1330. [Google Scholar] [CrossRef] [PubMed]
- Takashima, K.; Koga, Y.; Anzai, T.; Migita, K.; Yamaguchi, T.; Ishikawa, A.; Sakashita, S.; Yasunaga, M.; Yano, T. Evaluation of Fluorescence Intensity and Antitumor Effect Using Real-Time Imaging in Photoimmunotherapy. Pharmaceuticals 2022, 15, 223. [Google Scholar] [CrossRef] [PubMed]
- Takashima, H.; Koga, Y.; Tsumura, R.; Yasunaga, M.; Tsuchiya, M.; Inoue, T.; Negishi, E.; Harada, M.; Yoshida, S.; Matsumura, Y. Reinforcement of antitumor effect of micelles containing anticancer drugs by binding of an anti-tissue factor antibody without direct cytocidal effects. Journal of Controlled Release 2020, 323, 138–150. [Google Scholar] [CrossRef] [PubMed]
- Chen, Y.K.; Pleskow, D.K. SpyGlass single-operator peroral cholangiopancreatoscopy system for the diagnosis and therapy of bile-duct disorders: A clinical feasibility study (with video). Gastrointestinal Endoscopy 2007, 65, 832–841. [Google Scholar] [CrossRef] [PubMed]
|
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).