3.1. Generator Response Behavior
In comparing the excitation voltage of the function generator to the resulting coil current, it is evident that the system of the induction generator and the resonant circuit system exhibits a sluggish response at the chosen pulse frequencies of 7 and 14 Hz, preventing it from reproducing the input signal even remotely. The damping present in the oscillating circuit of the induction generator, coupled with its internal circuit topology, causes the current measured in the induction coil to lag behind the input signal. This behavior is observed for both rectangular and sinusoidal excitation voltages, as depicted in
Figure 3 for a frequency of 7 Hz.
When comparing the two curves of coil current, it is evident that a sinusoidal excitation voltage exhibits a smoother profile compared to a rectangular excitation voltage with steep edges. At the falling edge of the rectangular signal, the generator also manages to achieve a rapid decrease in output power. Conversely, during the rising edge, a very gradual increase in output power is observed. This implies that the induction generator system is capable of reacting much faster to a decrease in power than to an increase. Furthermore, the measurement results suggest that at control frequencies above 10 Hz combined with larger amplitudes, the maximum specified power cannot be achieved by the generator before the excitation voltage starts to decrease again. Additionally, it becomes apparent that, on the one hand, a smaller difference between the high and low levels of the excitation signal, and on the other hand, a high frequency of the excitation oscillation lead to an increasingly indistinguishable base and pulse current. For both types of excitation voltage, the oscillation of the coil current is asymmetric, so that the negative amplitude is significantly higher than the positive amplitude. The response behaviour of the induction generator could not be analysed over the whole excitation voltage range from 0 to 10 V. The reason for this is that the change of current (RMS-value) over time exceeds the threshold of the current probe at an excitation voltage of 7 V (i.e. 70 % of the generator power) or more at the given frequency.
3.2. Droplet Formation and Detachments
The use of a stopped-down lens during recording, combined with the short exposure time resulting from the high frame rate, limits the visible field to the glowing wire and reflections, rendering the surrounding environment black in the images. Despite the insights that can be derived from the data captured by the camera, the partial occlusion of the wire tip by the induction coil restricts the possibilities for detailed analysis. Thus, depending on the selected process parameters, the area of the wire where droplet detachment occurs may be obscured.
Figure 4 illustrates the camera view and the resulting shot of the wire.
The recorded data demonstrates that the high frame rate of the camera is sufficient to capture the process in slow motion, including the wire feed movement, melting of the wire, droplet formation, and detachment. When the technology is operated within an appropriate process window, the droplet formation is highly regular and the resulting droplets are uniform in size and shape. Outside the favorable process window, comparatively larger drops form in some cases and detach from the wire end at irregular intervals. In addition, the movement of the melt at the end of the wire increases and spatter formation becomes more intense.
Figure 5 exemplifies a sequence of images corresponding to a favorable set of parameters (60-70 % generator power, 7 Hz pulse frequency), wherein a regular and uniform droplet detachment is observed. The moments of droplet detachment exhibit temporal scattering of less than 10 ms. Furthermore, the observed droplet sizes appear consistent, although their assessment is somewhat limited due to the low dynamic range in the capture area and the reduced resolution resulting from the close focusing distance of the installed lens and the necessary cropping. The location at which the molten wire end is constricted and the droplets detach is approximately at the height of the lower coil winding. At this generator power level, the droplet detachment frequency is 14 Hz.
The experiments with constant power, however, demonstrated that uniform droplet detachment can occur even without power pulsation. The detachment frequency increases with increasing generator power
up to a certain power level, denoted as
(approximately 65 % in these experiments). For
<
, lower droplet detachment frequencies are observed compared to
=
. Beyond this critical point, increasing the generator power
>
no longer leads to an increase in detachment frequency. However, the detachment process becomes more turbulent, and the molten flow intensifies. Since the detachment frequency remains constant, the residence time of the droplets at the wire end is the same as at the lower power level
. Consequently, they experience increased heating, resulting in a comparatively higher temperature at the moment of detachment. This is clearly recognizable by the brighter glow of the droplets. An exemplary droplet detachment for
>
is depicted in
Figure 6. The droplet size and the detachment process itself are comparable to those described earlier in
Figure 5.
Figure 7 displays the sequence of images from an experiment with pulsed generator power where the average power
theoretically lies at
=
. However, due to the previously described inertia of the generator, it practically falls below
. As a result, a substantially lower droplet detachment frequency of approximately 2 Hz and an irregular melting behavior are observed. The temporal scattering of the detachments at around 300 ms is significantly higher than in the previously described case, indicating that droplet formation is comparatively more stochastic. Due to the lower power level, the region where the wire constricts and the droplet detaches is located well below the induction coil, likely triggered primarily by the gravitational force of the droplet. The energy supplied to the continuously fed wire is not sufficient to fully melt and detach it within the induction coil. Furthermore, the comparatively weaker luminosity of the droplets suggests that a lower droplet temperature is present compared to those shown in
Figure 5 and
Figure 6. Overall, the detachment process is more random, with the free end of the wire pinching at two points, and the molten droplets exhibiting more movement. It is conceivable that in this case, the power pulsation may have a rather adverse effect on the regularity.
Figure 8 shows a sequence of images depicting a slow droplet formation and detachment at
<
. Although the material of the fed wire can be melted, the power is insufficient for rapid detachment. As a consequence, large-volume droplets form and detach from the wire end only at a late stage and below the induction coil. The droplet detachment frequency at this generator power level is 4.7 Hz. Compared to the droplets in
Figure 7, it can be assumed that there is a higher temperature and thus a higher average power. There is only one point of constriction, and the temporal spread of detachments is less than 30 ms, thus significantly lower.
The analysis of the camera recordings shows that the frequencies of droplet detachment, both in constant power and pulsed operation, exceed the practically achievable pulse frequency of the generator technology used. The power decreases rapidly on the falling edge, but the power increases too slowly on the rising edge. To beneficially influence the process of droplet detachment through pulsed induction coil current, the authors see the need for adjustments in the circuit topology of the induction generator. With reliable power pulsation at frequencies of at least 15 Hz, a beneficial influence on the process can be expected.
3.3. Size distribution of drops
Based on a visual analysis with the naked eye, it is evident that the detached droplets possess an almost spherical shape. Consequently, it can be inferred that the described method allows for sufficient determination and comparability of droplet sizes and their volumes.
Figure 9 provides a comparison between the captured photograph and the evaluation performed using ImageJ.
In the first step, particles were colored red based on a threshold value derived from the grayscale of all particles, thus separating them from the white background. In the second step, the actual particle analysis was performed using a filter, which excluded smaller particles such as spatter or chippings of oxide layers, thereby preventing them from distorting the analysis results.
In
Figure 10, the size distribution during operation with constant power is depicted. In comparison,
Figure 11 and
Figure 12 illustrate the size distribution of droplets under pulsed power conditions at 7 and 14 Hz, respectively. The size analysis of the droplets shows that their projected area is generally between 3.5 and 5.5 mm
2. Based on the assumption of spherical geometry, this corresponds to droplet diameters between 2.11 and 2.65 mm and indicates that the deviation of the actual absolute droplet size is relatively small. When comparing the experiments conducted at constant power, it is evident that at 65 % power as opposed to 75 %, a greater number of outliers are observed in the histogram, resulting in a wider spread. These outliers predominantly consist of larger droplets exceeding the average size. With higher power, the droplets tend to exhibit a slight reduction in size, which can be attributed to a marginally higher detachment frequency due to the constant wire feed rate. However, the data appears to display a relatively improved normal distribution at lower power settings.
By pulsing the power sinusoidally at a frequency of 7 Hz around 65 % (62-68 %), the droplet distribution shifts towards larger droplets and exhibits an increased spread compared to constant power. This can be attributed, on the one hand, to the mismatch between the pulsing frequency and the actual droplet detachment frequency under constant power, as the pulsing process tends to hinder rather than support the detachment process. On the other hand, as previously mentioned, pulsing results in the average power being slightly below the theoretical value of 65 %. Furthermore, this effect is further amplified when the power amplitude increases (60-70 %), resulting in a rightward shift of the distribution curve within the histogram. When the average power is increased in pulse operation (70-80 %), the resulting droplets are detached at a faster rate, leading to smaller sizes. However, the pulsing frequency still opposes the actual droplet detachment frequency. As a consequence, the distribution in the histogram shifts to the left towards smaller droplets, while simultaneously exhibiting an increased spread.
This behavior changes when operating at a pulsing frequency of 14 Hz, which is closer to the actual droplet detachment frequency. Pulsing around the average power of 65 % (60-70 %) does not exhibit a significantly negative impact on the droplet size distribution and spread. While the power pulsing does not enhance droplet detachment, it no longer poses a notable disruption, as observed at 7 Hz. Increasing the average generator power to 75 % (60-90 %) further reduces droplet sizes, as they detach more rapidly from the wire end. However, the resulting spread is smaller compared to 7 Hz, as the pulsed power aligns closer to the actual detachment frequency. If the chosen power amplitude is too large (50-90 %), the inertia of the induction generator becomes more pronounced, as it can no longer produce a smooth output signal. Consequently, the droplet formation process is significantly disrupted. As a result, the spread increases, and the distribution shifts overall to the right, indicating larger droplet sizes.
3.4. Droplet Temperature and Efficiency
Figure 13 presents the power of the system recorded by the power meter in comparison to the measured droplet power.
Figure 14 depicts the average droplet temperatures and the calculated efficiency of each experiment. The measurements from the experiments with constant generator power demonstrate that the device’s power consumption exhibits a nearly linear relationship with the set power. It becomes evident that when operating in pulse mode with oscillation around a mean value, the power intake is lower compared to operating at a constant power level equivalent to that mean value. This can be attributed to the generator’s characteristic of quickly dropping to a lower power level but slowly rising back to a higher level, as discussed earlier. Consequently, the power consumption in pulse mode remains slightly lower than in constant mode, regardless of the pulse frequency. If the base power is increased while reducing the pulse power to the same extent, thereby keeping the average power constant, the power consumption of the system still increases due to the described behavior.
The material calculations using JMatPro revealed that the steel wire used has a solidus temperature of approximately 1460 °C and a liquidus temperature of 1517 °C. The calculated droplet temperatures indicate that they are mainly above these values, suggesting that the material collected by the calorimeter was completely melted. However, in the experiment with 55 % constant power, the temperature of 1293 °C is below the solidus temperature, yet the generated droplets still exhibit a spherical geometry. This suggests that there is an offset in the measurements, and the actual droplet temperatures are higher than calculated. A significant reason for this behavior is the measurement inaccuracy of the used Seebeck calorimeter, as demonstrated by [
29], which is approximately 9 %. This is particularly attributable to the long measurement time compared to the relatively short time of energy input. Nevertheless, the determined values can be compared to each other. The droplet temperatures correlate with the power consumption in such a way that the droplet temperature is higher when the generator’s power consumption increases. Overheating of the droplets during detachment through increased pulse currents for improved detachment is only partially possible, and the effect is relatively weak. With a faster response of the generator to increasing power input, such behavior could potentially be enforced.
In terms of efficiency, the process seems to fall short of expectations. While the overall efficiency of induction heating is generally considered relatively high compared to other heating processes, even an induction coil, as used here, which heats in the inner field, has a better inductor efficiency than, for example, a surface inductor that acts through its outer field [
11,
33,
34]. However, the calculated overall efficiency, considering the power consumption to thermal power in the droplets across all experiments, is only between 9 and 14 %. As previously explicated, given that the ascertained droplet temperature is expected to be lower than the actual value, it follows that the derived efficiency is somewhat diminished compared to the real magnitude. Therefore, future investigations will include an additional power measurement directly at the induction coil to better contextualize these results.