Preprint
Article

Looking into the Quantification of Forensic Samples with Real-Time PCR

Altmetrics

Downloads

100

Views

53

Comments

0

A peer-reviewed article of this preprint also exists.

Submitted:

08 May 2024

Posted:

09 May 2024

You are already at the latest version

Alerts
Abstract
The quantification of human DNA extracts from forensic samples plays a key role in the forensic genetics process. Quantification guarantees maximum efficiency and avoids repeated analyses, over amplified samples, or unnecessary examinations. Here, we present our experience using an extremely sensitive real-time-based method with the Quantifiler® Trio DNA quantification kit, which we have explored in the ISO/IEC17025 accredited test validation process conducted in our laboratory. In this study, our focus was on establishing a minimum threshold beyond which DNA typing is unnecessary, by defining the correlation of negative quantification results with the absence of genetic profiles resulting from DNA typing. We investigated the results in the range in which the technique is used (50-0.005 ng/μl). In the validation process of the internally accredited method, we used data from the commercial quantification system, based on the average values of the standard curve using a wide range of samples, including artificially degraded samples. This method was extensively used continuously for nine years to quantify the DNA extract coming from a wide range of samples from criminal cases. The validation allowed us to verify the limits of the technique, especially in the use of the minimum quantities necessary to obtain genetic profiles acceptable for sending to the Italian DNA database. This optimization of the protocol for analysing forensic samples allows us to optimise the protocol for analysing forensic samples. However, the technique is used in our laboratory with a completely manual preparative phase, and we have decided to investigate the contribution of human error more accurately. Thus, in our experience, we have found that it is preferable to use quantification ranges rather than exact limits before deciding how to analyse the extracts via PCR. In addition, here we present some preliminary data related to the analysis of certain samples that initially appeared negative during quantification, applying a concentration method for maximum yield from forensic samples of judicial interest. Our aim is to ensure that no source of evidence is lost in the reconstruction of a criminal event.
Keywords: 
Subject: Biology and Life Sciences  -   Biology and Biotechnology

1. Introduction

The identification of biological traces through DNA testing is now a well-established technique worldwide. It forms the basis for the construction of DNA databases, allowing for the comparison of traces, unknown individuals, and unidentified bodies through the implementation of genetic profiles generated with the highest quality standards [1]. The various phases of the techniques used are optimised within each laboratory, as there is no universally certified method issued by a public entity. Therefore, the reference standard is ISO/IEC 17025, which generally pertains to testing laboratories. In this work, we aim to revisit in a detailed manner the phase through which the qualitative and quantitative presence of human DNA in DNA extracts is verified. The importance of this analytical phase is underscored by the early guidelines of the forensic community. Already in 1992 one of the first recommendations of the International Society of Forensic Haemogenetics indicated that DNA Quantification of DNA is important for improving the quality of PCR-based DNA typing results. Specific working ranges of human DNA to serve as a PCR template are recommended. Therefore, initial quantification of human DNA should be performed [2]. Actually, in real casework the assumption that a trace may contain human biological material is sometimes contradicted by reality, as it is often possible to identify biological traces at crime scenes that are not of human origin. When it’s possible the spectrophotometric measurement method was chosen for the following reasons: it was simple, quick, very cheap and sufficiently reliable [3]. Especially when samples are taken after a long time after deposition, it is possible to find DNA of bacterial origin which is co-extracted together with human DNA. It is thus preferable to use real time PCR methods as 260 nm spectrophotometry is not suitable for discriminating the different types of DNA. Over the years, commercially available qPCR kits, such as PowerQuant™ System (Promega, USA) [4], QuantiFiler™ Trio (Thermo Fisher Scientific, Waltham, MA, USA) [5] or Investigator Quantiplex Pro (QIAGEN, Hilden, Germany) [6], were made available by commercial companies. These systems use TaqManR probes labelled with two fluorescent dyes: a reporter dye at the 5′-end and a quencher dye at the 3′end [7]. Briefly, when the probe is intact the two dyes are in close proximity due to the small size of the probe and the fluorescence of the reporter dye is suppressed. During the extension stage of a qPCR, the 5′-exonuclease activity of DNA polymerase displaces the bound TaqMan probes, and the reporter dye attached to the 5′-end is separated from the quencher dye at the 3′-end. The separation means the energy transfer that suppressed fluorescence when the two fluorophores were in close proximity is no longer occurring and the reporter molecule can now fluoresce. The amount of fluorescence detected correlates directly with the amount of human DNA present in the reaction, as the greater the amount of DNA, the more reporter dye molecules are cleaved, and the greater the fluorescence detected. Typically, as the qPCR program progresses, and the DNA increases exponentially, an amplification curve is generated using these fluorescence data, which (at the end of the qPCR program) are then used to calculate a DNA concentration [8]. Measured DNA concentrations are used to normalise the amount of DNA template to an optimal range for STR amplification reactions, ensuring STR profiles are in the linear range of the capillary electrophoresis instrument with minimal artefacts and stochastic effects. Every commercially available qPCR assays routinely used to determine the concentration of total human DNA as well human male DNA; they are based on the construction of standard curves with DNA at known concentrations. Each point of the curves is constructed using mean values ​​of serial dilutions assayed in duplicate.
Our experience from 2016 to today is based on use of Quantifiler™ Trio DNA Quantification Kit (ThermoFisher Cat. No. 4482910), for quantification of DNA extract from stains [9]. The Quantifiler™ Trio DNA Quantification Kit uses multiple-copy target loci for improved detection sensitivity and combines four 5′ nuclease assays. The human-specific target loci (Small Autosomal, Large Autosomal, and Y-chromosome targets) each consist of multiple copies dispersed on various autosomal chromosomes (Small Autosomal and Large Autosomal), or multiple copies on the Y-chromosome. The two separate target-specific human assays, one with a short PCR amplicon and one with a long PCR amplicon is able to determine the state of degradation of the sample. The fourth target (internal PCR control IPC) is a synthetic probe capable of identifying the presence of inhibitors in extracts. This quantification method is part of the analytical process of the laboratory's ISO/IEC17025 accredited method identified with denomination “DNA typing for human identification, mixed stains, Y-STR, paternity and kinship testing (genetic profile)” with the Italian Accreditation Body, Accredia (Lab nr. 1268) [10]. In the indicated period it was used for 240 experiments for a total of more than 10000 forensic samples. The amount of DNA in these samples can vary from tens of nanograms to a few picograms, corresponding to a few diploid genomes, or it can yield negative results in the absence of human genetic material.
The method we use consists in a “standard protocol” that is used when sufficient DNA is present to obtain genetic profiles from a single PCR reaction and in a so-called “low template scenario” protocol when the amount of template DNA is low. In this case a double amplification of the sample is performed and the results of the analysis is therefore derived from the comparison of the two replicates as described [11]. If the quantification result is below a certain limit, the PCR reaction is not carried out and the outcome of the assessment is that the sample does not contain sufficient human DNA. With the initial validation process we identified quantitative limits for applying the standard and low template scenario protocols. A visual representation of this protocol is represented in Figure 1.
Obviously, the analyses with low template scenarios have several limitations. Indeed, allele dropout, stutter, false alleles, and the influence of sporadic background contamination should be considered when interpreting these DNA profiles [12]. However, these profiles are often found on items touched, which are those most frequently requested by investigators for inclusion in the DNA database. Sometimes the only connection to a crime can be represented by the discovery of such a faint trace. For this reason, it is crucial to pay the utmost attention to extracting as much information as possible from these specimens. However, this activity must be balanced with a robust laboratory routine, identifying an optimal cost-benefit synthesis.
The quantitative limits mentioned above were determined through an initial laboratory validation. In DNA quantification, precision is crucial to ensure that the obtained results are reliable and consistent. Precision refers to the consistency and reproducibility of measurements obtained from repeated analyses of the same sample under the same conditions. Accuracy refers to the closeness of a measured value to the true or accepted value. However, we are aware that certain parameters can influence the final analytical results. One parameter that seems not to be evaluated with due attention is the uncertainty due to the use of measuring instruments and manual activities. This parameter is carefully evaluated in many other fields [13] and should be carefully assessed even within the forensic context, as suggested by some authors [14]. Indeed, an important factor regarding the reproducibility of the standard curve is the mixing function during pipetting of the serial dilution by the liquid handling system. In the method used in our laboratory, quantification requires all manual steps, including the construction of the calibration curve with the dilution of the control DNA and pipetting in the sample plate. In DNA quantification, accuracy is essential to ensure that the reported DNA concentration reflects the actual amount of DNA present in the sample. ISO/IEC 17025 mandates that accredited laboratories validate the accuracy of their quantification methods by comparing their results with reference materials of known DNA concentration. Thus, the piston pipettes we used are annually calibrated by an external supplier accredited for calibrations according to Accredia. The bioanalytical approaches used by the supplier involve gravimetry according to ISO 8655 guidelines (ISO 8655-6) [15]. However, all these precautions can be affected by human random errors in pipetting of standards and samples. Constructing the standard curve with a replicate of each control could compensate for pipetting uncertainty. Using the numerous data available, we have carried out a careful review of the individual points of the construction of the calibration lines, to verify how the revision of the trend of the calibration lines influences the final results. Based on the results obtained, appropriate actions were implemented to optimise this step in the forensic genetic method we use for DNA typing. This analysis can be considered a revalidation of the method, in accordance with point 7.2.2.2 of the ISO 17025:2015 standard [16].
Finally, we show that, even if the results of the quantifications of DNA extracts from minimal traces are practically zero, it is possible to use a sample concentration approach which in some cases has proven useful for generating genetic profiles of the trace donors. These indications suggest ample room for improvement for a technique that already provides fundamental results in the forensic genetics analysis process used in daily routine.

2. Materials and Methods

2.1. DNA Extraction

For automated purification of DNA from forensic casework and human samples, EZ1® Advanced XL instruments in combination with EZ1 Investigator kit were used (Qiagen, Hilde, Germany). The 40 μl final volume in TE was always used to obtain extracts at the highest possible concentration.

2.2. Analytical Phase of the Quantification Method

The human DNA used to generate the DNA quantification standards dilution series consists of pooled human male genomic DNA. Standards dilution series with the concentrations ranging from 50 ng/μl (Std. 1) to 0.005 ng/μl, or 5 pg/μl (Std. 5). When 2 μl of a sample at the lowest concentration (5 pg/μl) is loaded in a reaction, the well contains approximately 1.5 diploid human genome equivalents [17]. Table 1 shows the serial dilution of the control DNA used in double in each experiment. In each experiment, 2 µL of each sample to be dosed were used in single experiments in 96-well plates.

2.3. Validation of the Kit

The initial validation of the real time system was carried out by analysing the control DNA available with the kit. Ten-fold dilution series with 5 concentration points as known in Table 1 were used. The standard curve was carried out using serial two-fold dilutions and quantitative evaluations carried out using mean values. For each point of the curve the same dilutions of the standards were assayed, using them as if they were samples. For each point on the curve, the examination was conducted with five replicates. The degradation index value for these samples was also determined, calculated as the ratio between the concentration of small autosomal DNA and the concentration of large autosomal DNA.

2.4. Reanalysis of the QuantiFiler™ Trio

20 different experiments to construct the standard curve were assessed to establish the minimum and maximum values for Slope, Y-intercept, and R2 for each of the probes used. The threshold cycles of the internal IPC control were also observed to determine if the extraction method used ensured sufficient purification of the traces.
For an accurate validation of the system applicable to the internal ISO/IEC 17025 method, 757 samples from 20 independent experiments conducted in the period 2016-2024 were used. These samples all came from real forensic cases and included traces of blood, saliva, semen, epithelial cells, bone fragments, and tissues.

2.4.1. Standard Versus Sample/Sample Versus Standard

To verify if there were differences between the two replicates used in each of the twenty experiments for constructing the standard curves, the standards were reciprocally used as standards or samples. This was made possible because each experiment could be reanalysed using Design & Analysis Software ver. 2.8.0 (Thermo Fisher Scientific™, USA). We designated each of the two standards as "Sample 1" and "Sample 2" when the other replicate was used as the standard. To assess the differences between the two analyses, we employed the Bland-Altman plot [18]. The Bland-Altman plot is used to assess the measurement difference between two different measurement techniques. In fact, if the measurement of differences is normally distributed and the mean difference is statistically different from 0 (i.e., mean ± confidence interval does not contain the zero), then we can assert that the two measurement techniques are statistically different. The normality of the dataset was performed using the Kolmogorov-Smirnov test calculator [19].

2.5. Concentration of Apparently Negative Samples

In a preliminary experiment, ten DNA extracts from forensic samples taken from subungual grooves of known individuals that had yielded results <0.003 ng/μl in real-time quantification were concentrated using the Savant SpeedVac™ Vacuum instrument (Thermo Fisher Scientific™, USA). The samples were dried at room temperature and then diluted in TE buffer to 5 μl, suspending them for 30 minutes at 37°C. This way, the DNA extracts were concentrated nearly eightfold compared to the original extracts. No further quantification was performed on these samples.

2.6. Amplification

For the amplification phase, we use both the GlobalfilerTM PCR Amplification kit and the PowerPlex® Fusion 6C, with a maximum volume of 7.5 ul of extracted DNA. When determining the Y chromosome profile, especially in cases of sexual assault, we utilise the kit PowerPlex® Y23. Therefore, at least theoretically, the minimum quantity for an analysis with a low template scenario is at least 22.5 pg/PCR, and for an analysis with a standard protocol, it is 225 pg/PCR. Finally, the concentrated samples were fully amplified using the GlobalFilerTM kit under standard conditions.

3. Results

3.1. Extraction Method

Figure 2 shows the IP value graph for twenty experiments containing 757 forensic samples of various types. The values are generally distributed closely together and rarely exceed cycle threshold 29, observed in the standard sample showing the highest inhibition (50 ng/μl). Complete inhibition was observed in only two samples out of the entire set of forensic samples examined (2/757 undetermined). The quality of the extracts evaluated on a large number of samples showed a reduced quantity of inhibitors in a few samples (2/757). This is largely due to the technology adopted for DNA extraction. The magnetic particle technology, which combines the speed and efficiency of silica-based DNA purification, is fully automated and has effectively ensured the complete removal of inhibitors from the treated samples. The particles are ultimately separated from the lysates using a magnet, and even if some particles remain in the tube, they do not act as inhibitors. Conversely, the DNA is then efficiently washed and eluted in the TE buffer, ensuring the applicability of subsequent steps.

3.2. Performance of the Kit

Table 2 shows the results obtained from the analysis of the five replicates of each standard. The average values ​​of the five measurements were calculated, along with their standard deviations and the differences from the theoretical dosage of each standard. For the higher concentration points of the curve (50 ng/µl), the large probe returns values ​​that are slightly different from the expected (-0.998%), while the small and Y probes overestimate significantly (+20% and +21%). When dosing highly concentrated samples, it is therefore this probe that provides the most reliable results. From the next point (5 ng/µL) onwards, the large probe proves to be less efficient than the others, overestimating the actual sample (+7.99%). For values at the lower points of the curve (0.5, 0.005, and 0.005 ng/µL), there is good correspondence between the theoretical and experimentally determined values.
Figure 3 graphically depicts the above results, with detailed information available in the supplementary material of the article.
Figure 4 shows the results for slope, Y-interceptor, R2. The examination of the experimental parameters highlighted the following. A slope close to -3.3 indicates optimal, 100% PCR amplification efficiency. In the revalidation of the system, the examination of twenty different experiments regarding the standard curves showed an average slope value of -3.3 for the small and Y probes, and -3.4 for the large probe. The maximum values observed in individual experiments were -3.68 for both small and large probes, while the minimum was -3.15 for a single experiment for the small probe. The Y-intercept value indicates the expected CT value for a sample with Qty = 1 (for example, 1 ng/µL). Each laboratory should experimentally evaluate the Y-intercept values over time. In addition to variations that can be caused by pipetting of standards or minor lot-to-lot variations in the kits, Y-intercept can also be affected by and/or instrument-to-instrument variation. As expected, the Y-intercept for the large autosomal target is typically lower (mean value 23.0) than the Y-intercept for the small autosomal target (25.5) or the Y target (25.1). This is because of the higher copy number of the large autosomal target relative to the copy number of the small autosomal and Y targets. An R2 value ≥0.99 indicates a close fit between the standard curve regression line and the individual CT data points of quantification standard reaction. In the twenty different experiments the mean values for the three probes were always >0.99% (SA = 0.994, LA=0.997, Y=0.996) (Figure 4). As shown in the figure, in two experiments, we observed an R2 <0.98, which was actually due to different outcomes in the preparation of their respective standard pairs. From our experience, this parameter can indeed indicate a significant difference in the quantification results of the samples used for standard preparation.

3.3. Standard Versus Sample/Sample Versus Standard Results

We plotted the standard measurements on the x-axis and the relative difference on the y-axis. We calculated the mean of the relative differences, which was found to be slightly below 0 for all measures but heavily dependent on the very high values. In this paper, we are interested in analysing what happens with low concentration values. Therefore, we repeated the Bland-Altman plot for values below the threshold of 0.05 ng/µl. In this second approach, we found a non-significant difference between the two results (Sample 1 and Sample 2). To be more accurate, we also performed a normality test to verify the normality of the distribution of the relative differences (the y-axis values). As we can see in Figure 5, the measures of Sample 1 for concentrations above 50 ng/µl are lower than those of Sample 2. In fact, if we consider all the concentrations, we obtain that Sample 1 measures 7/9 compared to Sample 2. However, as shown in Figure 6, if we restrict ourselves to lower concentrations, we find that the two measurements are not significantly different.
So, even though using two standards at the same concentration might theoretically decrease pipetting differences, these measurements don’t show significant discrepancies (at least for value lower than 0.05 ng/µL). Therefore, we observe in Figure 6 the different behaviour of the differences for large or small quantity values. Consequently, we calculated the mean and confidence interval of the latter (e 95% confidence interval is calculated as mean ± 1.96 *Standard Error). Without censoring any value, we find that the confidence interval of the mean, slightly, does not include the zero, suggesting that the two measurement techniques are different. However, focusing on values within our range of interest (values below 0.05ng/µl) and calculating the mean and confidence interval, we obtain a relative difference of 0.02 ± 0.02, thus 0 is included, and therefore, no statistically significant differences exist between the two approaches within the range of 0 ng/µl to 0.05 ng/µl.
For both analyses, we assessed the normality of the dataset, Figure 7 (using the Kolmogorov-Smirnov test calculator and both resulted in distributions that are almost normal, allowing us to use the Bland-Altman method.

3.4. Make One Last Attempt

DNA extracts from samples that would not have been further examined were instead PCR amplified after being concentrated approximately 8-fold. Figure 8 shows the results of the electropherograms, demonstrating the presence of amplification products in samples with theoretical concentrations of 0.002 and 0.001 ng/μl. Comparison with the genetic profile of the donors confirmed the correct typing, with the presence of drop-in and drop-out as expected (Figure 9). A sample with a concentration of 0.000 ng/μl did not yield any typing results, except for a single allele at D3S1358. Although these are preliminary results from a few samples, this method indicates concrete possibilities for successfully determining genetic profiles even from samples containing very few cells which may not be detected by the quantitative test.

4. Discussion

Our service primarily focuses on genetic diagnoses in the prenatal field. However, due to the lack of law enforcement laboratories in the Tuscany region (Central Italy), we established a forensic genetics section in 2009, with an ISO/IEC 17025 accredited test since 2012. The laboratory serves as a reference point for forensic genetics for the Tuscany region [20]. As of today, there are another eight highly specialised Italian laboratories, not belonging to law enforcement, located in other Italian regions, to produce genetic profiles accredited according to ISO standards. The purpose of our service is to strive for the highest possible outcome for genetic profile determinations from items related to crimes. The ISO/IEC 17025 accreditation issued by the Italian national accreditation body, Accredia, indeed allows our laboratory to submit genetic profiles to the National DNA Database, managed by the Ministry of the Interior in Rome.
The accredited method requires a critical and constant review of the results obtained in the execution of the test. We are aware that to achieve accuracy in DNA quantification, sources of systematic error and measurement distortion, sources of uncertainty, must be minimised. Using real-time methodology, we can ensure the detection of human DNA, the identification of male components, and the presence of any inhibitors that interfere with the PCR phase.
Precision and accuracy are therefore critical aspects of DNA quantification in forensic science. We are aware that in the use of manual methods, especially in the pipetting operations of standards and samples to be quantified, random errors can occur. It is known that real-time PCR tests are extremely sensitive, and the detection of CT values ​​>35 may indicate the presence of extremely low quantities of DNA. In fact, it is possible to detect CT values ​​<40 for extraction control samples and negative control samples while performing a real-time PCR reaction with the QuantiFiler™ Trio DNA quantification kit. The detection of such a low quantity of DNA can vary from amplification to amplification based on stochastic effects. Therefore, it is recommended that each laboratory establish a CT value above which a positive result represents only a background signal [Quantifiler]. From our experience, we have verified the utility of conducting a thorough re-evaluation of the results obtained from numerous real-time experiments. The QuantiFiler™ Trio DNA quantification kit can detect DNA concentrations <5 pg/μl; however, at concentrations <5 pg/μl, stochastic effects or the statistical effect of random sampling of alleles present in a very low number of copies can produce significant variability in assay results. When using samples containing DNA in this concentration range, replicate analyses are indeed necessary to confirm the true absence of DNA. In the review of this phase of the process, it was possible to define more accurate limits based on the quantification results for subsequent choices regarding the possible determination of genetic profiles. This limit has been recalculated considering that the review of the quantifications performed by the laboratory showed that for the lower limits, there was about a 10% difference between measurements made with two different calibration curves. This discrepancy had already been observed in the validation process in which serial dilutions of control DNA had been measured as samples. Although the relative Bland-Altman in our area of interest did not show significantly relevant differences, with an average ratio of 0.02±0.02 (<(x1-x2)/(x1+x2)/2>=0,02±0.02), meaning that on average one of the two measurements differs within a range of approximately 0% to 4% from the other. Therefore, as a precautionary measure, we have decided for the time being to adjust the limits based on the results of the quantification for our method. This allows us to make better choices by optimizing the conduct of the test, pending a more effective implementation that could involve the introduction of an automatic method for the preparation of quantification plates, which could certainly reduce random errors. The vast majority of laboratories use manual methods to quantify forensic extracts, and thus this work should help to pay attention to the execution of this phase. We believe that the critical review of quantifications reported in this study can contribute to an improvement in the entire forensic analysis process, emphasising how to take into account the uncertainty always present in any laboratory activity and, ultimately, the consistency of human activities. Figure 10 shows the modification in the quantification. Highlights in yellow the new limits adopted in our working protocol in light of this study.
However, there are even more substantial improvements that we plan to implement shortly. In fact, the use of extract concentration, when their concentration is below <0.003 pg/μl applied to some samples obtained from scrapings under the nails of unknown donors, has demonstrated the possibility of generating genetic profiles suitable for comparisons. The concept of usability stems directly from the indications of the Italian DNA Database, for which a profile can be included for comparisons on a national level if at least seven genetic markers are typed, including amelogenin. Of course, this new sample analysis method will require a careful validation process, especially since the profiles were generated with a single PCR. A possible development could therefore involve, after observing quantities <0.002 ng/μl (new lower limit set by this study), a first PCR cycle, followed by verifying the quality of the generated genetic profile. If the result shows an acceptable profile, a second PCR could be performed to generate an additional genetic profile and thus determine a consensus profile. If the generated profile is not usable, the sample could be concentrated and then amplified from the extract suspended in a minimal amount of TE buffer. The method could also be further improved by using a reduced reaction volume, as demonstrated by some authors [21]. This, in addition to potentially enhancing the functionality of quantifying extracts from forensic samples, will also significantly reduce costs. As already mentioned, an additional improvement to this phase could be the introduction of automation, which helps reduce errors in pipetting standards during calibration curve construction and sample retrieval.

5. Conclusions

This study is based on a decade-long experience in forensic sample analysis at the laboratory designated by the Tuscany Region as a reference point for forensic genetics. The mandatory validation processes in the accreditation context have been useful for a critical review of any part of the test process. We aimed to revalidate the quantification phase, which precedes the decision on whether a trace extract should undergo typing analysis or not. This critical review has highlighted that the uncertainty due to the technical phases of quantification preparation requires redefining the limits for this decision-making process to avoid underestimating samples that may appear negative. Moreover, this study also emphasises another aspect that will require further, in-depth validation before being applied for the analysis of real samples. By applying a concentration step to some seemingly negative extracts and subjecting the sample to a single PCR phase, usable genetic profiles for comparisons were generated. It is therefore evident that this latter analytical phase must be included in the process before truly considering a trace as unsuitable for generating a genetic profile for identification purposes. It is evident that there are still opportunities to improve this critical analytical phase, which is perfectly aligned with the expectation of continuous improvement as indicated by the ISO/IEC 17025 standard.

Author Contributions

Conceptualisation, U.R. and D.C..; methodology, U.R. and D.C.; software, D.C..; validation, I.C., C.C., E.P., I.G., M.P. and U.R.; formal analysis, U.R. and D.C..; data curation, D.C..; writing—original draft preparation, U.R.; writing—review and editing, U.R.; supervision, U.R and D.C. All authors have read and agreed to the published version of the manuscript.

Funding

This work received no financial support.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

The data presented in this study are available on request from the corresponding author. The data are not publicly available due to privacy.

Acknowledgments

This study is the result of over ten years of work by a laboratory accredited according to the ISO17025 standard for forensic genetics testing and from the meticulous systematic review of experimental data conducted for the continuous improvement of activities for the judiciary.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Butler, J.M. Advanced Topics in Forensic DNA Typing: Methodology. Elsevier Science: Ed. Amsterdam, The Netherlands, 2014.
  2. Bär, W.; Brinkmann, B.; Lincoln, P.; Mayr, W.; Rossi, U.; Budowle. B; Eisenberg, A.; Fourney, R.; Gill, P.; Rand, S. Editorial: Recommendations of the DNA Commission of the International Society for Forensic Haemogenetics relating to the use of PCR-based polymorphisms. Forensic Sci. Int. 1992, 55, 1–3. [Google Scholar]
  3. Nicklas, J.A.; Buel, E. Quantification of DNA in forensic samples. Anal. Bioanal. Chem. 2003, 376, 1160–1167. [Google Scholar] [CrossRef] [PubMed]
  4. Ewing, M.M.; Thompson, J.M.; McLaren, R.S.; Purpero, V.M.; Thomas, K.J.; Dobrowski, P.A.; DeGroot, G.A.; Romsos, E.L.; Storts, D.R. Human DNA quantification and sample quality assessment: Developmental validation of the PowerQuant(®) system. Forensic Sci Int Genet. 2016, 23, 166–177. [Google Scholar] [CrossRef] [PubMed]
  5. Holt, A.; Wootton, S.C.; Mulero, J.J.; Brzoska, P.M.; Langit, E.; Green, R.L. Developmental validation of the Quantifiler(R) HP and Trio Kits for human DNA quantification in forensic samples. Forensic Sci. Int. Genet. 2016, 21, 145–157. [Google Scholar] [CrossRef] [PubMed]
  6. Vraneš, M.; Scherer, M.; Elliott, K. Development and validation of the InvestigatorR Quantiplex Pro Kit for qPCR-based examination of the quantity and quality of human DNA in forensic samples. Forensic Sci. Int. Genet. Suppl. Ser. 2017, 6, e518–e519. [Google Scholar] [CrossRef]
  7. Jothikumar, P.; Hill, V.; Narayanan, J. Design of FRET-TaqMan probes for multiplex real-time PCR using an internal positive control. Biotechniques 2009, 46, 519–524. [Google Scholar] [CrossRef]
  8. McDonald, C.; Taylor, D.; Linacre, A. PCR in Forensic Science: A Critical Review. Genes 2024, 15, 438. [Google Scholar] [CrossRef] [PubMed]
  9. Holt, A.; Wootton, S.C.; Mulero, J.J.; Brzoska, P.M.; Langit, E.; Green, R.L. Developmental validation of the Quantifiler® HP and Trio Kits for human DNA quantification in forensic samples. Forensic Science International: Genetics 2015, 28, 145–157. [Google Scholar] [CrossRef]
  10. https://services.accredia.it/accredia_labsearch.jsp?ID_LINK=1734&area=310&dipartimento=L&.
  11. Caragine, T; Mikulasovich, R; Tamariz, J; Bajda, E; Sebestyen, J; Baum, H; Prinz, M. Validation of testing and interpretation protocols for low template DNA samples using AmpFlSTR Identifiler. Croat Med J. 2009, 50, 250–267. [Google Scholar] [CrossRef] [PubMed]
  12. Gill, P.; Gusmão, L.; Haned, H.; Mayr, WR.; Morling, N.; Parson, W.; Prieto, L.; Prinz, M.; Schneider, H.; Schneider, P.M.; Weir, B.S. DNA commission of the International Society of Forensic Genetics: Recommendations on the evaluation of STR typing results that may include drop-out and/or drop-in using probabilistic methods. Forensic Science International: Genetics 2012, 6, 679–688. [Google Scholar] [CrossRef]
  13. Guan, X.L.; Chang, D.P.S.; Mok, Z.X.; Lee, B. Assessing variations in manual pipetting: An under-investigated requirement of good laboratory practice. Journal of Mass Spectrometry and Advances in the Clinical Lab 2023, 30, 25–39. [Google Scholar] [CrossRef]
  14. Schwenzer, R.; Schoneberg, A.; Pflug, W. Implementation of a robotized real-time PCR setup for the use of the QuantifilerTM Human DNA Quantification Kit. Forensic Science International: Genetics Supplement Series 2008, 1, 68–70. [Google Scholar]
  15. 8655-6, I. Piston-operated volumetric apparatus—Part 6: Gravimetric methods for the determination of measurement error. (ISO, 2022).
  16. Ricci, U. Establishment of an ISO 17025:2005 accredited forensic genetics laboratory in Italy. Accred Qual Assur 2014, 19, 289–299. [Google Scholar] [CrossRef]
  17. Quantifiler™ HP and Trio DNA Quantification Kits. USER GUIDE. Publication Number 4485354 rev. H.
  18. Bland, J.M.; Altman, D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1, 307–310. [Google Scholar] [CrossRef] [PubMed]
  19. Statistics Kingdom. (2017). Multiple Linear Regression Calculator. (October 30, 2023) [web application].https://www.statskingdom.com/410multi_linear_regression.html.
  20. Delibera n.1053 del 11-09-2023 - https://www301.regione.toscana.it/bancadati/atti/DettaglioAttiG.xml?codprat=2023DG00000001319.
  21. Yoonjung, C.; Hyo, S.K.; Min-Hee, K.; Moonhee, P.; Hansol, K.; Yang, H.L.; Dong, S.L. Validation of Reduced Reagent Volumes in the Implementation of the Quantifiler® Trio Quantification Kit. J Forensic Sci 2018, 63, 517–525. [Google Scholar]
Figure 1. Diagram representation of genetic DNA forensic process.
Figure 1. Diagram representation of genetic DNA forensic process.
Preprints 105996 g001
Figure 2. IP-value in real experiments from casework samples.
Figure 2. IP-value in real experiments from casework samples.
Preprints 105996 g002
Figure 3. Correlation between expected and observed quantification of DNA control.
Figure 3. Correlation between expected and observed quantification of DNA control.
Preprints 105996 g003
Figure 4. Slope, Y-interceptor, R2 in twenty different experiments.
Figure 4. Slope, Y-interceptor, R2 in twenty different experiments.
Preprints 105996 g004
Figure 5. The relative Bland-Altman plot.
Figure 5. The relative Bland-Altman plot.
Preprints 105996 g005
Figure 6. The relative Bland-Altman plot for quantity values lower than 0.05.
Figure 6. The relative Bland-Altman plot for quantity values lower than 0.05.
Preprints 105996 g006
Figure 7. Report of the normality test.
Figure 7. Report of the normality test.
Preprints 105996 g007
Figure 8. Raw data of concentrated DNA samples from casework.
Figure 8. Raw data of concentrated DNA samples from casework.
Preprints 105996 g008
Figure 9. Electroferograms for the same samples as in figure 8, compared to the reference DNA from a saliva sample of the donor (only the blue channel is shown). Although some "drop-in" and "drop-out" phenomena were observed in some loci, the donor's genetic profile is evident for almost all DNA markers.
Figure 9. Electroferograms for the same samples as in figure 8, compared to the reference DNA from a saliva sample of the donor (only the blue channel is shown). Although some "drop-in" and "drop-out" phenomena were observed in some loci, the donor's genetic profile is evident for almost all DNA markers.
Preprints 105996 g009
Figure 10. The forensic genetics process with modifications after the revalidation process.
Figure 10. The forensic genetics process with modifications after the revalidation process.
Preprints 105996 g010
Table 1. Standards dilution series.
Table 1. Standards dilution series.
Preprints 105996 i001
Table 2. Experiment with DNA control dosed like samples.
Table 2. Experiment with DNA control dosed like samples.
Preprints 105996 i002
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.
Prerpints.org logo

Preprints.org is a free preprint server supported by MDPI in Basel, Switzerland.

Subscribe

© 2024 MDPI (Basel, Switzerland) unless otherwise stated