The first solution (
Section 3.1) describes how remote powering is added to the two-wire bus without adding noise to the measure bioimpedances and without disrupting synchronization and communication between the sensors and the central unit.
Section 3.2 then addresses how remote powering can be made safe for medical regulations while significantly reducing the volume of the sensors (e.g., from 7.5 cm
3 to 0.3 cm
3) and the power consumption (e.g., from 5.8 mA to 150 µA), while additionally detailing circuit strategies for measurement, current injection, and management of return currents.
3.1. Legacy Approach with 500 Hz Powering and Off-the-Shelf Components
Figure 5 shows a modified version of the circuit in
Figure 2 with remote powering circuitry. In this scheme, the voltage source
of the central unit provides power to the 2-wire bus that is then “harvested” by the cooperative sensors. Harvesting power from the bus is depicted in
Figure 5 by a current source whose current is in phase with the voltage
(power is consumed when current and voltage are in phase).
It is important that the power supplied from the central unit does not disturb the measurements. Although the resistance of the bus wires may be low, the supply current is high, so the voltage drop on the lower wire will add to the voltage of interest,
. To avoid interference, the power supply frequency can be placed outside the frequency band of the measurements. The first order bandwidth of the bioimpedance is, for example, 40±0.5 kHz for 25 current channels and 40 frames per second. Therefore, powering could be done at 0 Hz. However, if the biopotentials are to be measured simultaneously by the same sensors, [
14] has shown that 0 Hz is not a good choice. The first solution described in [
14] proposes powering with a square wave at 500 Hz. Interference with the bioimpedance measurement must be expected at 500 Hz and its harmonics, i.e., 1 kHz, 1.5 kHz, 2 kHz, … If the frequency band for biopotential measurements is chosen at, for instance, 40±0.5 kHz, possible interferences could be expected at 39.5, 40.0, and 40.5 kHz. IQ demodulation will shift this band to −500, 0, and 500 Hz. The other harmonics of the powering will also be shifted, but will remain at frequencies 1 kHz, 1.5 kHz, 2 kHz, … When sampled at 1 kHz, all these spectral lines will create aliases at 0 and 500 Hz and therefore, an anti-aliasing filter must be used to reduce their power. A delta sigma analogue-to-digital converter as used for biopotentials in [
14] can efficiently avoid aliasing at 0 Hz, but there will still be some aliasing at 500 Hz. Moreover, at 0 Hz there will still be the original 40 kHz harmonic, which is small, however, because it is the 100th harmonic. Moreover, for most applications, the useful bioimpedance signal is in the respiration or cardiac bands that do not include 0 Hz, i.e., a high pass comb filter is used to remove the 0 Hz component of the bioimpedance signal (which is for 25 current channels at 0 Hz, 40 Hz, 80 Hz, …). The 500 Hz alias can be removed with a notch comb filter at 20 Hz (i.e., the Nyquist frequency of the bioimpedance channels) if the number of current channels is odd (e.g.,
) since 500 Hz is in the middle of the
bioimpedance spectra.
The harmonics generated by the 500 Hz power supply frequency are of low enough amplitude to avoid significant interference with the digital communication (the 2 Mb/s of [
18,
19] were modified to 1.28 Mb/s, in both directions, to conform to the frequency required by the delta-sigma converter to sample at 1 kHz). It was observed, however, that bits near the rising/falling edges of the 500 Hz power signal were disturbed, so these bits were removed from the communication payload. In our prototype, 110 bits were removed in total, reducing the throughput by 17%.
In
Figure 4 and [
18,
19], upstream communication (from the central unit to the sensors) is implemented as a voltage source and the downstream channel (from the sensors to the central unit) as current sources. Alternatively, the upstream and downstream channels could be interleaved through time domain multiplexing to allow both channels to use either voltages or currents to communicate. In
Figure 5, the current sources are implemented with the Thevenin equivalent voltage source with a resistance in series because practically this is easier to implement using the digital output pin of a microcontroller or FPGA. The capacitor, inductor, and resistor are chosen to create a first-order bandpass filter that filters the communication band from the biopotentials while simultaneously attenuating the high frequency harmonics from the digital signal. Note that
,
, and
are configured in parallel in both the central unit (voltage source
at 0) and in the sensors (the capacitance in series with the inductance is chosen so that its impedance is negligible at communication frequencies). The received signal is then measured as the voltage across the RLC circuit. The digital signal can then be reconstructed with a high-pass filter and Schmitt trigger before demodulation in the D block to obtain
.
A consequence of putting the cooperative sensors in parallel on the 2-wire bus is that there is a practical limit to the number of sensors that can be connected. The LC circuit attenuates current at the communication frequency such that the voltage on the RLC circuit is effectively a voltage divider consisting of the transmitter’s resistance and all other resistances of the receivers in parallel. Consequently, the received voltage is the transmitter voltage divided by the number of receiving units (i.e., sensors and the central unit). This approach practically limits the number of units to approximately 25.
The 500 Hz supply square wave is generated in the central unit by the voltage source , which can easily be implemented using switching transistors. Since the inductances of the central unit and sensor are tuned for the communication frequencies, they have a negligible impact on the supply current. The 500 Hz square wave can thus be rectified by diodes in the sensors to generate a positive and negative voltage rail across the storage capacitors.
In the implementation of
Figure 5, there is no floating power supply and therefore no bootstrap as in
Figure 4. The current source
must be designed with high output impedance. Moreover, its current returns via both upper and lower wires, thereby affecting the measurement of the electrode potential since the lower wire is used as reference potential.
3.2. Approach Addressing the Safety Issue with Powering at 1 MHz and ASIC
While the circuit in
Figure 5 reduces power supply interference with bioimpedance measurements and with communication on the 2-wire bus, it does not address safety for medical devices. In medical device standards, the allowed
patient leakage current at 500 Hz is 100 μA for
type bf devices [
15], which is already ten times higher than the allowed current at d.c. However, measuring a 100 μA leakage current by monitoring the power supply signal is a difficult task. Assuming a per sensor consumption of 8 mA (see next section) and 25 sensors, the supply current on the bus is 200 mA, requiring accurate measurement of 1 part in 2000. In addition, all sensors would require a current source buffer that could make closed-loop adjustments to ensure the current at each sensor is precisely 8 mA.
To better ensure compliance with the allowed patient leakage current, the power consumption of the sensors should be reduced (e.g., by a factor of 20 down to 400 μA) and the power supply frequency should be increased, e.g., to 1 MHz where the standards allow up to 10 mA of leakage current. These changes facilitate detecting excessive leakage current (1 part in 1). Note that the 8 mA and 400 μA mentioned above are the current of a sensor measured in the bus. The sensor itself consumes half of this current, i.e., 4 mA and 200 μA, respectively. The factor-of-two difference arises from the conservation of energy: the supply voltage is with a current of (rms value for square waves) which allows sensors with the dual half-wave rectifier to have a supply and current for the electronics.
To reduce power consumption by a factor of 20, we developed an ASIC (application-specific integrated circuit) that optimized each electronic function. In addition, we eliminated the digitization (analog-to-digital converter). The transmission of analogue values instead of bits has also increased the throughput (which allows considering other signals, in our case the sound picked up by a stethoscope embedded in the electrode). The 1 MHz power supply is now interleaved with the communication, i.e., in every two periods there is one for power supply and one for communication [
22].
Figure 6 depicts the design. The inductances are no longer needed, which is beneficial given the challenges of integrating inductors in silicon. The rectifier diodes allow the storage capacitors to be recharged (harvesting period) when the voltage
on the bus is high enough. When the voltage
is low enough for the diodes to be in blocked state, the communication current source of one sensor is used to transmit the information to the central unit. For the other sensors, the current source is disabled (i.e., no current is consumed by these sensors during the communication period). Therefore, the central unit decides if the period is a harvesting or communication period with the level of the voltage
. As shown in
Figure 7, by choosing if the harvesting period follows or precedes the communication period, the central unit can send a 0 or 1 bit to the sensors (Manchester code). This upstream communication channel may be used for instance to configure the sensors. Note that the Manchester code always has a transition (H to L or L to H) in the middle of its period (see blue edges in
Figure 7). Therefore, two harvesting periods in a row (H to H separated by a blue line) can easily be interpreted by the sensors as a synchronization marker rather than a 0 or a 1. From this marker, every sensor can count the number of edges and recognize its communication period defined by its ID. Note that this scheme is similar to that described in [
14] but simpler in the sense that the edges are always regular and can readily be used as clock signal (without the need of PLL or timer).
The regulated supply rails VCCF and GNDF (specific for each sensor) can be bootstrapped by using a follower to control the reference potential of the LDO voltage regulators. Assuming an LDO gain,
(i.e., the LDO outputs a current
where
is the voltage error of the LDO output), the input impedance of the open-loop circuit is magnified by
at low frequencies. High gain at low frequencies can be achieved if
(see
Figure 6) behaves like a capacitance at low frequency. At higher frequencies, for stability reasons, it is preferable for
to behave like a resistance. The open loop input impedance is essentially the input impedance of the follower (typically 10 pF). The bootstrap magnifies this impedance by
, allowing the circuit to have a very high input impedance at low frequencies [
22]. Compared to
Figure 5 where bootstrapping is not implemented, this bootstrap also makes shielding of the sensor input more efficient and natural, as the ground and power rail planes implicitly provide driven shielding. Additionally, the output impedance of the current source (
in
Figure 5) as seen from the body is magnified by the bootstrap. If
is an order of magnitude larger than the lower wire impedance (which is easily the case), the part of injected current
that is conducted by the lower wire is negligible. As a result, the bioimpedance measurement is not affected by the wire impedance.
Before transmission, the biopotential is filtered by the voltage divider
,
, amplified, and IQ demodulated. The filter is a band-pass filter centered at 50 kHz (EIT current frequency of the ASIC variant) that removes the electrode offset (up to 300 mV) and possible biopotentials (generally lower than 10 mV). A perfect IQ demodulation would not require such filtering since the multiplication of the signal by a cosine at 50 kHz would swap the bioimpedance and biopotential bands. But in practice it cannot be perfect, and the high-pass filter will prevent residual biopotential in the demodulated bioimpedance signal. Most of the bioimpedance signal is at or close to 0 Hz (typically 98–99%). However, the signal of interest is generally at breathing or cardiac frequencies. Therefore, compression of the low frequencies (with a filter inverse to a low-pass filter) is advisable before transmission. At reception by the central unit, the original signal can be decompressed with a low-pass filter. When the bioimpedance signal is the result of more than one current channel, such compression filter has also to be applied shifted at 40 Hz, 80 Hz, … (comb filter). This pre-processing improves the signal-to-noise ratio of the analog communication.
Figure 8 shows a possible implementation of the comb filter with transfer function
with
(z transform variable),
kHz (sample rate),
(number of current channels), and
where
Hz (corner frequency chosen here as the lowest respiration frequency) and
(compression factor of the bioimpedance signal at 0 Hz). The current sample voltage is applied over
and
in series, with the capacitance
having been shorted (switch
closed) just before. The charge accumulated by
and
in series is copied, thanks to the operational amplifier, in the capacitance
(also shorted just before). The resulting transfer function is therefore:
When the switch is open, the charges in the capacitance are held (sample and hold function) and the potential is ready for transmission. While switch is open and as soon as the voltage has been transmitted, switches and can be closed. They are reopened before another switch is closed for the next sample. The memorized value in is available 25 samples later.
The M modulator (see
Figure 6) selects the correct time slot for the sensor to communicate. All sensors are synchronized to sample their electrode potential simultaneously, and the measured value to be transmitted is stored in a capacitor until being transmitted. While this paper describes the measurement of bioimpedance, the developed ASIC can also measure biopotentials [
14] and, if interfaced with an electret, body sounds (stethoscope).