The Devices
In clinical practice, two BIA measuring methods have been primarily developed: the “external band electrode method” and the “implanted device lead method” [
10]. The prevailing approach for measuring total body compartments is the hand-to-foot external band method [
10] which was originally introduced by Hoffer and revised by Nyboer [
18,
19]. The approach aims to decrease the contact impedance between the skin and the electrodes and was validated or the first time by Lukaski on 140 healthy adults [
2]. In this pioneer study, the authors avoided electrode polarization and reduced the effects of skin impedance by employing four electrodes: one pair for passing the current through the body and the other to detect the resulting voltage [
20]. In detail, the electrodes are placed on the dorsal side of the hands and feet, with a distance of at least 3-5 cm between the signaling and detection electrodes [
20]. Tetra-polar measurements are taken on a supine subject for generally 15 minutes to allow the stabilization of the body fluids [
20]. Ideally, before placing the electrodes, the surface should be cleaned with alcohol to avoid measurement bias [
20].
Another way to utilize the external band electrode method is through thoracic BIA, which uses two pairs of external electrodes placed on the body surface on the thorax [
21]. This method exploits the phenomenon whereby variations in fluid volume and blood flow determine changes in the frequency of propagating waves rather than in the amplitude of the signal [
21,
22]. Therefore, detecting the phase shift of electrical currents (known as bioreactance) also allows the assessment of cardiac output and vascular resistance [
21,
22]. The method is also known as impedance cardiography (ICG) [
22].
A different method is based on the assessment of intrathoracic impedance by measuring the impedance between the device case, typically implanted in the left pectoral region, and the lead in the right ventricle [
23]. Impedance has long been used to check for lead integrity in pacemakers or defibrillator devices and intrathoracic bioimpedance measurements have been also developed and integrated into cardiac implantable electronic devices (CIED) to check volume status [
24]. In fact, changes in impedance can be determined across two relatively fixed points, thus minimizing distortion, or variations in electrode placement. Therefore, the primary purpose of implanted device–based BIA is to monitor over time the clinical status in chronically ill patients as an additional function of the implanted device [
24,
25].
Although BIA technology has been widely investigated, its application is, to this day, still limited due to many factors and far from being integrated in the clinical routine setting [
26].
In the following article we will refer generally to the term BIA for what concerns the bioimpedance analysis technology, nonetheless we will specify for each study the type of technique (e.g. MF-BIA, BIVA, BIS) that has been used.
Clinical Relevance of BIA in HF Patients
HF presents the highest rate of hospitalizations among 65-year-old patients [
27], with high morbidity and mortality, consuming vast resources of the public health system around the world [
28]. The projected absolute number of people suffering from HF is approximately 26 million worldwide, and it is considered a growing epidemic with increasing incidence and prevalence [
5,
29]. In patients with acute decompensated HF (ADHF), body fluid congestion, presenting different clinical degrees of dyspnea, orthopnea, fatigue, jugular venous distension, rales, and edema, was demonstrated to be associated with re-hospitalization and high mortality rate [
30]. Moreover, it is widely recognized that the pathophysiological determinant of the worsening of the patient’s state is caused by increased left ventricular filling pressures, dysregulation of the effective circulatory volume, and consequent extracellular fluids retention and mostly intravascular congestion [
31] that, if left untreated, leads to pulmonary edema [
32]. It is known that approximately 90% of patients admitted to the emergency department exhibit signs of fluid overload [
33], and as indicated by another study, about 50% of them are discharged with some degree of congestion [
34].
When patients are admitted to the hospital with excessive edema, intravenous diuretic therapy is commonly administered; however, since the monitoring is mostly performed through clinical observations, physical examination, or evaluation of the diuresis, the overall body volume assessment is not precise [
35,
36].
Determining diuretic therapy efficacy also depends on the patient’s diuretic resistance, which often makes the in-hospital assessment of the body volume challenging. These difficulties are also quite expected in home-setting conditions [
37].
In many peripheral wards, body weight changes are utilized to evaluate the performance of diuretic therapy, assuming that a deterioration of HF has been defined as an increase of 2 kg in 48/h [
38]. Nonetheless, different studies have shown that body weight monitoring presents a low sensitivity for the timely detection of congestive events and does not completely correlate with the process of a patient’s decompensation [
39,
40].
Body volume assessment is critical at home since preventing acute decompensation events is an unmet necessity [
41]. Therefore, many studies have investigated how to manage the volume status of HF patients by the time they leave the hospitals to predict and avoid re-hospitalizations [
38]. Although the cardioMEMS system, in which a small sensor is implanted the pulmonary artery to measure the pulmonary artery pressure, has been proven to be an effective method to prevent re-hospitalizations and reduce mortality in this set of patients, its use is limited by the invasiveness, costs, market penetration and potential complications [
42,
43,
44]. Non-invasive methods or devices, employed mostly for regularly checking the patient’s body weight and congestion symptoms, have been studied to identify timely a potential heart failure decompensation event [
45]. BIA has emerged to detect accurately the level of body fluids, and it has been studied as a potentially earlier predictor for decompensation events rather than an increase in body weight [
30].
A selection of the most relevant prospective clinical studies in HF patients studying hard endpoints that showed a beneficial or a neutral effect of BIA in both intra-hospital and homecare settings are summarized in
Table 1.
One of the earlier studies on BIA performed by Packer et al, also known as the Prospective Evaluation and Identification of Cardiac Decompensation by ICG Test (PREDICT), was conducted to assess the feasibility of ICG in predicting clinical deterioration in ambulatory patients with HF [
46]. In detail, the study prospectively evaluated 212 stable patients with chronic HF (CHF) who presented a recent episode of clinical decompensation; patients underwent clinical evaluation and blinded ICG testing every 2 weeks for 26 weeks and were followed up for the occurrence of death or hospitalization [
46]. The study combined clinical parameters such as NYHA class and blood pressure with ICG parameters such as velocity index, thoracic fluid content (TFC) index, and left ventricular ejection time, producing a composite score. The score showed to predict with good accuracy an HF event during the next 14 days of follow-up (p = 0.0002). Moreover, visits with a high-risk ICG score presented an 8.4% event rate during the next 14 days and predicted 41.6% of the events [
46].
In 2010 the group of Di Somma et al aimed to verify whether BIVA could be a valid methodology for the assessment of fluid overload in ADHF, if there was a correlation with BNP, and if BIVA could be used in the management of the diuretic therapy in acute settings [
47]. A total of 51 patients were enrolled, and their hydration state, BNP level, and caval index were evaluated at the following time points: at admission, after 24 and 72 hours, and at discharge. Moreover, a follow-up by phone was performed after 3 months. BIOVA measured higher values of hydration state in ADHF patients compared to controls. The most interesting result was that patients with an average hydration value >80.5% had a correlation with events at 3 months (death or rehospitalization for cardiogenic event) with a sensitivity of 22% and specificity of 94.2% [
47].
In 2017, a prospective, multicenter, observational study by Santarelli et al. investigated the prognostic role of quantitative reduction of congestion during hospitalization by using BIVA serial evaluations in patients admitted for ADHF [
48]. Both clinical and BIVA evaluations were performed at admission and discharge. A follow-up phone call was carried out at 90 days, and the primary endpoint was a composite of re-hospitalizations for HF or all-cause mortality. BIVA correctly predicted the primary endpoint at both admission (area under the curve (AUC) 0.56, p<0.04) and discharge (AUC 0.57, p<0.03). When combined with clinical evaluation, the prediction of the primary endpoint significantly increased for all-cause mortality or re-hospitalizations at 90 days (AUC 0.97, p<0.0001). Noteworthy, an increase of in-hospital resistance variation (dR/H) of more than 11 Ohm/m was associated with overall survival [
48]. No significant results were found about the in-hospital reactance variation (dXc/H), underlining the concept that R is primarily influenced by changes in hydration status [
48]. Furthermore, AHF patients discharged with the presence of at least one congestion sign were experiencing a lower survival [
48].
BIA external devices have also been studied to assess their potential utilization for remote management in HF patients. The MUSIC study (Multisensor Monitoring in Congestive Heart Failure) was conducted to develop and validate an algorithm for predicting impending acute HF decompensation using a multi-parameter approach, including BIA obtained from an external device adhered to the chest [
49]. This study found that the multisensor algorithms which included ICG as one of its sensors could accurately predict HF exacerbations with a sensitivity of 87% and a specificity of 87%. The algorithm also predicted HF aggravations an average of 6 days before clinical events occurred [
49]. Moreover, BIA-derived measures of TFC were sensitive for predicting HF decompensation 9 to 11 days in advance. The study found that a TFC increase of 7.5% from baseline could predict clinical events with a sensitivity of 77% and a specificity of 64% [
49].
Furthermore, in a prospective study by Gyllensten et al., 91 patients with CHF were monitored for an average of 10 months using a weight scale and a wearable BIA vest in home settings to predict HF exacerbations. One of the most important results was that BIA algorithms could better predict impending decompensation within 2 weeks compared to changes in weight (cross-validation estimate was 60% for BIA vs 33% for body weight). However, the authors noted that many alerts detected through the BIA device and its algorithms were not associated with clinically overt decompensation and related HF hospitalizations [
50].
The SENTINEL-HF study, instead, tested a wearable BIA device called a fluid accumulation vest. This device is noninvasive, capable of transmitting data via a mobile phone, and employs an automated algorithm to predict recurrent HF events. Among the study participants with sufficient data (n=57), an algorithm analyzing thoracic BIA showed 87% sensitivity (95% CI 82-92), 70% specificity (95% CI 68-72), and 72% accuracy (95% CI 70-74) in the identification of HF events [
51].
Different studies concentrated on analyzing intrathoracic congestion by applying BIA to the current generated from the pacing wires of pacemakers and defibrillators. One of the first studies on an ICD equipped with intrathoracic impedance monitoring was conducted by Wang et al. in a canine model [
23,
26,
52]. The study demonstrated that impedance to electrical flow on the lead of these devices can reveal changes in thoracic congestion, mainly due to impending HF [
23].
A study performed by Check-Man Yu et al. estimated that an implantable system capable of measuring intrathoracic impedance could identify fluid overload in patients before HF hospitalization and determine its correlation with standard measures of fluid status during hospitalization [
26]. In detail, the automated detection of decreases in intrathoracic impedance via pacemaker and implantable cardioverter-defibrillator devices alerted in advance of the beginning of decompensation due to volume overload, providing timely information for titration of the medication [
26]. Therefore, internal impedance-meter-equipped CIEDs have been developed with the product most known as Opti-Vol. In the DOT-HF trial, Opti-Vol affected the clinical outcomes of CHF patients, most likely due to additional data obtained from this method [
53]. In the study, the number of deaths was comparable (p=0.54), while the number of outpatient visits was higher in the access arm (250 versus 84; p<0.0001). Interestingly, the intervention was associated with a borderline statistically significant increase in the primary endpoint of all-cause mortality and HF hospitalizations. This increase was mostly due to increased HF-related admissions caused by the Opti-Vol method, which led to an increase in unneeded hospitalizations, thus not addressing the authors’ main hypothesis [
53].
Eventually, a recent multi-centric study performed by Anshory et al. investigated the role of a non-invasive method to assess hemodynamic parameters and total body congestion via BIA (NICaS)[
54]. In this work, the authors aimed to evaluate the above-mentioned technology in a cohort of Asian and Caucasian AHF patients predicting mortality and re-hospitalization for HF within 30 days. As a main result, data at admission significantly predicted 30-day cardiovascular mortality and rehospitalization. Moreover, at discharge, a value of cardiac output obtained using NICaS predicted significantly a 30-day rehospitalization. Noteworthy, total peripheral resistance index on admission and during 48–72 hours showed each an AUC > 0.70 in predicting mortality and rehospitalization compared to clinical congestion score and NT-proBNP. These findings underline the importance of the role of combined methodologies, such as total body water and hemodynamics, to better understand the patients’ clinical status.
Clinical Relevance of BIA in CKD Patients
Chronic kidney disease (CKD) is defined as a reduction in kidney function that persists for more than 3 months, with leading causes including diabetes and hypertension, as well as infectious glomerulonephritis, vasculitis, congenital anomalies of the kidney and urinary tract (CAKUT), inherited and autoimmune diseases. Regardless of the etiology, CKD is characterized by progressive and accelerating loss of renal function caused by self-maintaining processes of fibrosis and atrophy of kidney tissues. Throughout the years, CKD can lead to end-stage kidney disease (ESKD), leading to potential kidney transplant or dialysis. The prevalence of CKD in the USA is as high as 40% among people over 70 years, with most patients having CKD stage 3 (GFR 30-59 ml/min/1.73 m2) [
55]. The main clinical manifestations are caused by the accumulation of uremic toxins, anemia, metabolic acidosis, malnutrition, altered electrolytes, and impaired water homeostasis, leading to volume expansion, hypertension, and peripheral/pulmonary edema. Chronic fluid overload is experienced by 40-50% of hemodialysis (HD) patients [
55].
The volume status of a patient is a crucial parameter in the nephrology clinical practice, especially for those with concomitant renal and heart failure, in which both fluid overload and volume depletion events are frequent and are associated with CKD progression and poor outcomes [
56,
57].
A selection of the most relevant prospective clinical studies in CKD patients studying hard endpoints that showed a beneficial or neutral effect of BIA is presented in
Table 2.
It has become clear that BIA (i.e. BIS) is the most widely used method to evaluate volume status and guide fluid management in nephrological patients [
58,
59]. In HD, BIS is already employed to assess the overhydration state, enabling clinicians to evaluate the amount of fluid that ultrafiltration should remove [
58,
59].
Hur et al performed a well-structured randomized trial to investigate the effect of BIS-guided fluid management on cardiovascular outcomes in hemodialysis patients [
60]. The authors used time-averaged fluid overload (TAFO) as a more accurate variable of the fluid status of patients. TAFO, calculated on pre-dialysis fluid overload (FO) and interdialytic weight gain (IDWG) (TAFO = FOpre – IDWG/2) was used to increase or decrease post-dialytic weight. Authors showed that bioimpedance fluid management significantly reduced left ventricular mass index (LVMI) and improved systolic and diastolic blood pressure pre- and post-dialysis compared to standard care [
60].
Onofriescu et al showed in a randomized trial that a strict BIA-based (BIS) fluid management not only improved survival by reducing mortality significantly (Hazard ratio (HR) = 0.100 (95% CI, 0.013-0.805; p=0.03)), but also surrogated endpoints such as arterial stiffness, relative fluid overload and systolic BP improved [
61].
In another randomized trial, ABISAD-III researchers provided an algorithm to determine post-dialysis weight based on the evaluation of fluid status by Body Composition Monitor [
62]. Incidence of acute fluid overload (AFO), cardiovascular events, hypertension, or intradialytic complications were significantly reduced in the intervention group [
62].
Other works and specific metanalysis studies have confirmed that excessive hydration defined by bioimpedance is an independent predictor of mortality in patients with ESKD [
63,
64].
A recent detailed methodological review and meta-analysis study on the use of whole-body BIA in dialysis patients was conducted on 46 manuscripts on the ESKD population; these studies were sub-grouped according to whether they used phase angle (PA)/BIVA, normalized ECW (ECW/TBW) or the overhydration index (OHI) [
65]. The work evidenced that OHI is an independent predictor of mortality in ESKD patients independent of the influence of comorbidity and that different parameters, such as PA and OHI, act as similar predictors of outcome [
65].
Another meta-analysis investigated the impact of bioimpedance-guided fluid management on two separated subsets of patients: CKD stage 3-5 and ESKD dialysis-dependent patients [
66]. To our knowledge, this is the largest meta-analysis published on BIA-based monitoring of fluid management in dialysis patients. The authors analyzed hard and secondary outcomes, and they showed a significantly reduced all-cause mortality (HR = 0.64, 95% CI: 0.41, 0.99) and lower blood pressure with the use of BIA [
66].
A systematic review and meta-analysis conducted by the NHS on more than 5,000 patients showed that BIA may lower overhydration in HD patients and improve blood pressure control, but it did not definitively prove a reduction in overall mortality [
67]. The authors suggested that the difference could be due to the new studies in the meta-analysis and the increased number of peritoneal dialysis (PD) patients. Indeed, patients with residual renal function usually have a lower risk of mortality. Thus, the inclusion of more studies with PD may have led to a significant difference [
67]. Differently, another recent study failed to demonstrate an improvement in short term clinical outcomes [
68]. Nevertheless, results were not discordant in finding a significant reduction in hospitalization rate and an impressive, although not statistically significant, trend in the improvement of the primary composite endpoint consisting of death, acute myocardial infarction, cerebral infarction, cerebral hemorrhage, and peripheral vascular disease (HR = 0.487, 95% CI 0.217-1.091, P = 0.08) [
68,
69].
Diagnosis, Therapy, and Risk Stratification in HF and CKD
BIA measurements can provide valuable information to clinicians for the diagnosis, treatment, and risk stratification of HF and CKD patients.
Figure 1A,B show a visual representation of the most relevant clinical benefits of BIA for HF and CKD patients, respectively.
ADHF = acute decompensated heart failure; BIA = bioimpedance analysis; CKD = chronic kidney disease; HF = heart failure; LVMI = left ventricular mass index
Various studies have repeatedly shown that BIA aids in the diagnostic assessment of the hydration status of HF and CKD patients.
Velazquez-Cecena et al. aimed to evaluate the correlation of low-, intermediate-, and high-risk groups for ADHF as determined by ICG parameters with LVEDP and serum BNP. The study’s main outcome was that patients considered at high risk for ADHF as determined by ICG presented significantly higher levels of LVEDP and BNP compared to lower-risk groups [
70].
A study by Parrinello et al was designed to evaluate the diagnostic role of segmental and whole-BIA in the differential diagnosis of ADHF from the other causes of acute dyspnea across a broad population presenting to the emergency department [
71]. One of the study’s main results revealed that whole body and segmental BIA were strong predictors of ADHF alone or in combination with BNP at the multiple regression analysis (AUC 0.989). Moreover, ROC curves were significantly higher (P < 0.0001) for segmental BIA (AUC = 0.963; CI: 0.935 - 0.982) and whole-body BIA (AUC = 0.934; CI: 0.902 - 0.961) than for LVEF (AUC = 0.814; CI: 0.764 - 0.857) and Framingham score (AUC = 0.681; CI: 0.625 - 0.734) [
71].
In a study of ambulatory patients with HF by Gastelurrutia et al., BIVA was utilized instead to successfully differentiate between stable and unstable HF; stable patients were found to have a significantly lower impedance-measured fluid load ratio and lower nt-proBNP than the unstable ones [
72].
Many studies concentrated on the role of BIA to guide and assess the efficacy of diuretic therapy in ADHF.
In a study by Zink et al., segmental BIA measurement was employed for the in-hospital monitoring of cardiac re-compensation in HF patients, which revealed the location of fluid accumulation and shift within the body during diuretic treatment [
73].
In the study by Di Somma et al., the authors demonstrated a direct correlation between BIVA and urine output at 72 hours, showing how the efficacy of diuretic therapy and the effectiveness of these measurements go hand in hand with the normalization of BNP and caval index values at discharge. Thus, the authors suggested that BIVA could be more useful than clinical signs in managing diuretic therapy in ADHF patients [
47].
Concerning CKD, a study aimed to evaluate the clinical usefulness of BIA in assessing volume status in patients receiving maintenance dialysis. The results indicated that BIA can accurately assess fluid volume status in patients receiving maintenance dialysis as measured by ECW and it can predict volume overload and hypertension in patients with end-stage renal disease. Lastly, it pointed out that BIA can be used to monitor changes in volume status over time and to guide clinical decision-making in managing fluid volume in patients receiving maintenance dialysis [
74].
The group of Khan et al. demonstrated the efficacy of BIS in the stratification of CKD patients based on their volemic status and therefore randomized them to different diuretic therapy approaches, with good results compared to controls [
58].
Detecting body fluids congestion in HF is challenging and many patients are discharged from the hospitals in a still congested state [
75,
76]. In a study on hospitalized ADHF patients, Massari et al demonstrated that BIVA could predict the total length of hospital stay independently. In other words, the more congested the patients were, the longer the staying in the hospital [
76].
Other studies investigated the role of BIA (i.e. BIVA) and BNP or NTproBNP for risk stratification of HF patients [
47,
77,
78]. In the study by Massari et al, BIVA was more accurate than BNP in the detection of peripheral congestion in both ADHF (AUC was 0.88 vs. 0.57 respectively; P < 0.001) and in chronic HF (AUC was 0.89 vs. 0.68, respectively; P < 0.001). Algorithms that include BIVA together with NTproBNP or BNP showed to be extremely accurate for patients´ prognosis assessment [
77].
Eventually, a metanalysis on the role of BIA (leg bioimpedance) in relation to incident HF in the general population, with more than 50.0000 patients, was presented by Lindholm et al. in 2018.[
79] This multivariable model that included leg bioimpedance, age, sex, and self-reported prior myocardial infarction showed good discrimination for future HF hospitalization, highlighting the potential benefits of adopting this simple, non-invasive, and cost-effective measure [
79].
Future Perspectives
Although the clinical relevance of BIA has been repeatedly demonstrated in the past years, its everyday utilization in the clinical wards or for remote monitoring remains limited.
One of the main problems regarding BIA adoption is the lack of large, randomized trials that could justify a complete reimbursement of the solution. Further studies should question the benefits of BIA, comparing the scenarios in which it is employed against those in which it is not applied, with the hypothesis that it can improve the current standard of care. This holds true for in-hospital use, but it is even more relevant for home settings.
Another point that needs to be addressed when considering its adoption is the complete integration of the measurement of HS into the clinical work routine.
Currently, the assessment of HS in the hospital is still mostly based on physical examination by the medical doctor, while patients use only weight scales at home. Therefore, the detection of HS should be accessible not only at the patient’s bedside but also at the patient’s home.
Hydration status should be an easily visualizable and actionable parameter, similar to blood pressure, blood glucose, or temperature, and channelled into a clinical decision tree.
Thus, it is crucial to have dedicated and instructed personnel collecting and interpreting the information. However, future solutions should focus on automatizing the information workflow and introducing smart sensors.
A recent study showed that a handheld bioimpedance system for the assessment of body fluid status could be used to assist the patient in home-based ESKD treatments [
83]. This device consisted of a custom-made handheld tetrapolar bioimpedance spectrometer and a textile-based electrode garment for total body fluid assessment [
83]. Future solutions should follow those developments.
We believe optimizing processes and improving clinical outcomes should drive HS adoption, with patients’ needs as the focus of our attention.
Another extremely relevant aspect concerns the creation of personalized pathology parameters that, through AI and the analysis of large datasets, should guide the personalization of treatments. Indeed, combining different biomarkers with BIA has been shown to be very promising for prognostic purposes. For this reason, our study group recently finalized a study currently under revision to assess the role of BIA, biomarkers, and other vital parameters to create a risk prediction algorithm that could detect mortality and re-hospitalizations within 6 months.